首页> 外文期刊>Physics in medicine and biology. >Application of an M-line-based backprojected filtration algorithm to triple-cone-beam helical CT.
【24h】

Application of an M-line-based backprojected filtration algorithm to triple-cone-beam helical CT.

机译:基于M线的反投影滤波算法在三锥束螺旋CT中的应用。

获取原文
获取原文并翻译 | 示例
           

摘要

To increase the scanning speed of cone-beam CT, a scanning configuration with three pairs of x-ray sources and panel detectors was proposed. In this paper, we apply the M-line-based backprojected filtration algorithm to this scanning configuration and propose a fast implementation for this algorithm. Compared with existing implementations, the advantages of our implementation are as follows: (1) it can directly reconstruct horizontal slices (i.e. the slices perpendicular to the rotation axis), without 3D interpolation; (2) it does not need to calculate the backprojection intervals for any point; (3) it has a high degree of parallelism and is suitable for acceleration on graphic processing units or other parallel systems.
机译:为了提高锥形束CT的扫描速度,提出了具有三对X射线源和面板探测器的扫描配置。在本文中,我们将基于M线的反投影过滤算法应用于此扫描配置,并提出了该算法的快速实现。与现有的实施方式相比,我们的实施方式的优点如下:(1)无需3D插值就可以直接重建水平切片(即垂直于旋转轴的切片)。 (2)不需要计算任何一点的反投影间隔; (3)具有很高的并行度,适合在图形处理单元或其他并行系统上加速。

著录项