Value of continuous glucose monitoring for minimizing severe hypoglycemia during tight glycemic control.

机译:在严格的血糖控制过程中,连续血糖监测对于最大程度降低严重低血糖的价值。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: : Tight glycemic control can potentially reduce morbidity and mortality in the intensive care unit but increases the risk of hypoglycemia. The most effective means to avoid hypoglycemia is to obtain frequent blood glucose samples, but this increases the burden to nursing staff. The objective of this study was to assess the ability of a real-time continuous glucose monitor to reduce hypoglycemia (blood glucose <60 mg/dL [3.3 mmol/L]) during standard care or tight glycemic control effected with a proportional integral derivative insulin titration algorithm. DESIGN: : Real-time continuous glucose monitor profiles obtained from an ongoing prospective randomized trial of tight glycemic control were retrospectively analyzed to determine whether the continuous glucose measure had prevented instances of hypoglycemia. SETTING: : Cardiac intensive care unit. PATIENTS: : Children 3 yrs of age or younger undergoing cardiac surgery were studied. INTERVENTIONS: : Intravenous insulin infusion and rescue glucose guided by the real-time continuous glucose monitor and the proportional integral derivative algorithm in the tight glycemic control arm (n = 155; target glucose 80-110 mg/dL [4.4-6.1 mmol/L]) and the real-time continuous glucose monitor in the standard care arm (n = 156). MEASUREMENTS AND MAIN RESULTS: : No reduction in hypoglycemia was observed with real-time continuous glucose monitor alarms set at 60 mg/dL (3.3 mmol/L) (zero of 19 occurrences of blood glucose <60 mg/dL [3.3 mmol/L] detected); 18 of 40 subsequent incidences of hypoglycemia were detected after the alarm threshold was increased to 70 mg/dL (3.9 mmol/L). In the tight glycemic control arm, eight incidences were reduced in duration and an additional eight events were prevented with intravenous glucose. In the standard care arm, three of nine occurrences of hypoglycemia were detected with the duration reduced in all cases. On average, one to two false hypoglycemia alarms were observed in each patient. CONCLUSIONS: : The real-time continuous glucose monitor in combination with proportional integral derivative control can reduce hypoglycemia during tight glycemic control. The real-time continuous glucose monitor can also reduce hypoglycemia during standard care. However, false alarms increase the overall nursing workload.
机译:目的:严格的血糖控制可能会降低重症监护病房的发病率和死亡率,但会增加低血糖的风险。避免低血糖的最有效方法是频繁获取血糖样本,但这增加了护理人员的负担。这项研究的目的是评估在标准护理或采用比例积分衍生物胰岛素进行严格的血糖控制期间,实时连续血糖监测仪降低低血糖(血糖<60 mg / dL [3.3 mmol / L])的能力。滴定算法。设计:回顾性分析正在进行中的严格血糖控制的一项前瞻性随机试验中获得的实时连续血糖监测资料,以确定连续血糖测量是否可以预防低血糖症。地点:心脏重症监护室。患者:研究了3岁或以下接受心脏手术的儿童。干预措施:在严格的血糖控制组中,实时实时葡萄糖监测仪和比例积分微分算法指导下静脉输注胰岛素和挽救葡萄糖(n = 155;目标葡萄糖80-110 mg / dL [4.4-6.1 mmol / L ])和标准护理臂中的实时连续葡萄糖监测器(n = 156)。测量和主要结果:设置为60 mg / dL(3.3 mmol / L)的实时连续血糖监测仪警报,未观察到低血糖症的减少(19次血糖<60 mg / dL [3.3 mmol / L]中的零发生]检测到);将警报阈值提高到70 mg / dL(3.9 mmol / L)后,随后检测到40次低血糖事件中的18次。在严格的血糖控制臂中,减少了八次发作的持续时间,并通过静脉内葡萄糖预防了另外八次发作。在标准护理组中,检测到9种低血糖发生中的3种,并且在所有情况下持续时间都缩短了。平均而言,每位患者观察到一到两个虚假的低血糖警报。结论:实时连续血糖监测仪与比例积分微分控制相结合可以降低严格血糖控制期间的低血糖。实时连续血糖监测仪还可以减少标准护理期间的低血糖症。但是,错误警报会增加总体护理工作量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号