Drug administration via high flow nasal cannula (HFNC) has been described in pediatrics but the amount of albuterol delivery with an HFNC is not known. The purpose of this study is to quantify aerosol delivery with heliox and oxygen (O(2)) in a model of pediatric ventilation. A vibrating mesh nebulizer (Aeroneb Solo, Aerogen) was placed on the inspiratory inlet of a heated humidifier and heated wire circuit attached to a pediatric nasal cannula (Optiflow, Fisher & Paykel). Breathing parameters were tidal volume (V(t)) 100 ml, respiratory rate (RR) 20/min, and I-time of 1 sec. Albuterol sulfate (2.5 mg/3 ml) was administered through a pediatric HFNC with O(2) (100%) and heliox (80/20% mixture). A total of 12 runs, using O(2) and heliox were conducted at 3 and 6 L/min (n = 3). Drug was collected on an absolute filter, eluted and measured using spectrophotometry. The percent inhaled dose (mean +/- SD) was similar with heliox and O(2) at 3 L/min (11.41 +/- 1.54 and 10.65 +/- 0.51, respectively; P = 0.465). However at 6 L/min drug deposition was >/= 2-fold greater with heliox (5.42 +/- 0.54) than O(2) (1.95 +/- 0.50; P = 0.01). Using a pediatric model of HFNC, reducing delivered flow from 6 to 3 L/min increased inhaled albuterol delivery >/= 2-fold but eliminated the increase in inhaled drug efficiency associated with heliox.
机译:在小儿科中已经描述了通过高流量鼻导管(HFNC)进行药物给药,但是用HFNC输注沙丁胺醇的量尚不清楚。这项研究的目的是在小儿通气模型中量化与氦氧混合气和氧气(O(2))的气溶胶输送。将振动筛网雾化器(Aeroneb Solo,Aerogen)放在加热的加湿器的吸气口上,并将加热的导线回路连接到小儿鼻插管(Optiflow,Fisher&Paykel)。呼吸参数为潮气量(V(t))100毫升,呼吸频率(RR)20 /分钟和I次为1秒。硫酸沙丁胺醇(2.5 mg / 3 ml)通过带有O(2)(100%)和氦氧混合气(80/20%混合物)的小儿HFNC给药。使用O(2)和Heliox进行的总共12次运行以3和6 L / min(n = 3)进行。用绝对滤器收集药物,洗脱并使用分光光度法测量。吸入剂量的百分数(平均+/- SD)与氦氧混合气和O(2)在3 L / min时相似(分别为11.41 +/- 1.54和10.65 +/- 0.51; P = 0.465)。但是,在6 L / min的剂量下,Heliox(5.42 +/- 0.54)的药物沉积比O(2)(1.95 +/- 0.50; P = 0.01)大2倍。使用HFNC的儿科模型,将输送流量从6 L / min减少到3 L / min,可使吸入的沙丁胺醇输送> / = 2倍,但消除了与日光有关的吸入药物效率的提高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号