Objective The purpose of our study was to determine if the rate of lead-related complications was increased with the Medtronic CapSureFix MRI? SureScan? 5086 MRI pacing lead (5086; Medtronic Inc., Minneapolis, MN, USA) compared to the previous generation of Medtronic CapSureFix Novus? 5076 pacing lead (5076). Background The 5086 lead is a newly introduced active-fixation pacemaker lead designed to be used conditionally in a magnetic resonance (MR) scanner. This lead has specific design changes compared to the previous generation of 5076 pacing leads. Methods This study was a retrospective case control study of 65 consecutive patients implanted with two 5086 leads compared to 92 consecutive control patients implanted with two 5076 leads over a 14-month period at a high-volume tertiary care hospital. Results Pericarditis, pericardial effusion, cardiac tamponade, or death within 30 days of implant were seen in eight patients from the 5086 cohort and two from the 5076 cohort (odds ratio 6.3, 95% confidence interval 1.3-30.8, P = 0.02). Lead dislodgement occurred in four of the 5086 patients and in none of the 5076 patients (P < 0.03). Conclusions In a high-volume center, the incidence of pericarditis, cardiac perforation, tamponade, death, and lead dislodgement was significantly higher with the MR-conditional Medtronic 5086 lead when compared to the previous generation Medtronic 5076 lead.
机译:目的我们的研究目的是确定Medtronic CapSureFix MRI是否能增加铅相关并发症的发生率? SureScan? 5086 MRI起搏导线(5086; Medtronic Inc.,美国明尼苏达州明尼阿波利斯)与上一代Medtronic CapSureFix Novus? 5076起搏导线(5076)。背景技术5086导线是新近推出的主动固定起搏器导线,旨在有条件地用于磁共振(MR)扫描仪中。与上一代的5076起搏导线相比,该导线具有特定的设计更改。方法这项研究是一项回顾性病例对照研究,对65例连续植入2根5086根导线的患者进行了对比,而92例连续对照在14个月内植入2根5076根导线的对照患者。结果植入后30天内,心包炎,心包积液,心脏压塞或死亡发生在8例5086队列和2例5076队列(赔率6.3,95%置信区间1.3-30.8,P = 0.02)。 5086例患者中有4例发生铅移位,而5076例患者中无一例发生铅移位(P <0.03)。结论在一个大容量的中心,与上一代Medtronic 5076铅相比,MR条件的Medtronic 5086铅使心包炎,心脏穿孔,填塞,死亡和铅移位的发生率显着更高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号