首页> 外文会议>Annual Review of Progress in Applied Computational Electromagnetics >Induced Current Analysis to Determine Magnetic Resonance Imaging (MRI) Safety of Cardiac Pace Maker Leads
【24h】

Induced Current Analysis to Determine Magnetic Resonance Imaging (MRI) Safety of Cardiac Pace Maker Leads

机译:诱导电流分析,以确定心脏步伐制造商的磁共振成像(MRI)安全性

获取原文

摘要

The heating potential due to induced currents in pace maker leads are analyzed by modeling the MRI environment and calculating the induced currents on three different lead geometries at two locations within a conducting medium. Peak SAR occurs near the bare electrodes attached to the leads even though currents are maximum in middle of portion of the two lead geometries analyzed.
机译:通过在导电介质内的两个位置计算三种不同引线几何的诱导电流来分析引起的步伐制造机引线引起的加热势。即使在分析的两个引线几何形状的中间的电流中,峰SAR在接合到引线附近的裸电极靠近。