Blood damage is an unavoidable side effect of extracorporeal circulation. The effects of blood damage on patients' hematocrit and erythropoietin requirement as well as other potential side effects have not been studied for uneventful treatments. Comparing long nocturnal dialysis with regular 4-hour, three times per week dialysis allows for the conclusion that the influence of blood damage caused by extracorporeal circulation is small compared with biochemical effects. Acute hemolysis is one of the few remaining mechanical problems of dialysis. Acute hemolysis is caused by obstructions within the extracorporeal circuit caused by manufacturing errors, kinking of blood tubing or user errors, or by a combination of excessive flow and improper cannula or catheter dimensions. The risk of acute hemolysis can be further reduced by industrial quality control, better design of dialysis equipment, and hemodialysis machine control. Adverse effects caused by chronic mechanical hemolysis need to be studied.
机译:血液损害是体外循环不可避免的副作用。血液损伤对患者血细胞比容和促红细胞生成素需求的影响以及其他潜在的副作用尚未针对平稳治疗进行过研究。将长时间的夜间透析与每周定期进行4小时,每周3次的透析进行比较,得出的结论是,与生化作用相比,体外循环引起的血液损害的影响较小。急性溶血是透析中剩余的少数机械问题之一。急性溶血是由体外回路内的阻塞引起的,这些阻塞是由制造错误,血液管道扭结或用户错误引起的,或者是由于流量过大和套管或导管尺寸不当而引起的。可以通过工业质量控制,更好的透析设备设计和血液透析机控制来进一步降低急性溶血的风险。由慢性机械性溶血引起的不良反应有待研究。