Analytical quality goals derived from the total deviation from patients' homeostatic set points, with a margin for analytical errors.

机译:分析质量目标源自与患者体内稳态设定点的总偏差,并存在分析误差。

获取原文
获取原文并翻译 | 示例
           

摘要

The deviation of test results from patients' homeostatic set points in steady-state conditions may complicate interpretation of the results and the comparison of results with clinical decision limits. In this study the total deviation from the homeostatic set point is defined as the maximum absolute deviation for 95% of measurements, and we present analytical quality requirements that prevent analytical error from increasing this deviation to more than about 12% above the value caused by biology alone. These quality requirements are: 1) The stable systematic error should be approximately 0, and 2) a systematic error that will be detected by the control program with 90% probability, should not be larger than half the value of the combined analytical and intra-individual standard deviation. As a result, when the most common control rules are used, the analytical standard deviation may be up to 0.15 times the intra-individual standard deviation. Analytical improvements beyond these requirements have little impact on the interpretability of measurement results.
机译:在稳态条件下,测试结果与患者体内稳态设定点的偏差可能会使结果的解释以及将结果与临床决策极限的比较变得复杂。在这项研究中,自稳态设定点的总偏差定义为95%测量值的最大绝对偏差,并且我们提出了分析质量要求,以防止分析误差将此偏差增加到比生物学引起的值高出约12%以上单独。这些质量要求是:1)稳定的系统误差应大约为0,并且2)控制程序将以90%的概率检测到系统误差,且该误差不应大于分析和内部分析结果总和的一半。个别标准差。结果,当使用最通用的控制规则时,分析标准偏差可能高达内部标准偏差的0.15倍。超出这些要求的分析改进对测量结果的可解释性影响很小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号