Contrast perception in digitized panoramic radiographs compared with their film-based origin.

机译:数字全景X射线照片的对比感知与基于胶片的原始对比。

获取原文
获取原文并翻译 | 示例
           

摘要

Objectives. We sought to compare the visual image quality of film-based and digitized panoramic radiographs through use of a hole-containing test wedge. Study Design. An aluminum wedge containing 100 cells, of which 90 were given shallow holes, was exposed in the film-based Orthophos CD panoramic unit. Two radiographs subjectively exhibiting optimum contrast were selected. Films were digitized with a charge-coupled device flatbed scanner at 300 dpi. Films and digitized images were rated cellwise by 2 similar groups of 50 observers each with respect to spot perception. Results. The mean sensitivity was 0.26 +/- 0.09 for film and 0.20 +/- 0.07 for digitized images (P =.000), with a pronounced decline in the latter in regions of high background density. The average specificity was 0.93 +/- 0.07 for film versus 0.92 +/- 0.08 for digitized images (P = 0.213). Conclusion. Film yielded a significantly higher sensitivity, but this absolute difference was actually small compared with that of the digitized images.
机译:目标。我们试图通过使用含孔的测试楔来比较基于胶片和数字化全景X射线照片的视觉图像质量。学习规划。在胶片型Orthophos CD全景单元中露出了一个包含100个单元的铝楔,其中90个具有浅孔。选择了两个主观表现出最佳对比度的射线照片。使用电荷耦合器件平板扫描仪以300 dpi将胶片数字化。胶片和数字化图像由2个类似的小组(每组50个观察者)进行逐细胞分级,每个小组关于点感知。结果。胶片的平均感光度为0.26 +/- 0.09,数字化图像的平均感光度为0.20 +/- 0.07(P = .000),后者在高背景密度区域明显下降。胶片的平均特异性为0.93 +/- 0.07,而数字化图像的平均特异性为0.92 +/- 0.08(P = 0.213)。结论。胶片产生了明显更高的感光度,但是与数字化图像相比,该绝对差实际上很小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号