Comparative localized linear accuracy of small-field cone-beam CT and multislice CT for alveolar bone measurements.

机译:小区域锥形束CT和多层CT在牙槽骨测量中的比较局部线性精度。

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVES: To compare the accuracy of cone-beam computerized tomography (CBCT) and multislice CT (MSCT) for linear jaw bone measurements. STUDY DESIGN: An ex vivo formalin-fixed human maxilla was imaged with both CBCT (Accuitomo 3D; Morita, Kyoto, Japan) and MSCT (4-slice Somatom VolumeZoom and 16-slice Somatom Sensation 16; Siemens, Erlangen, Germany). The MSCT images were reconstructed using different reconstruction filters to optimize bone visualization (U70u and U90u for VolumeZoom, H30s and H60s for Sensation 16). Before scanning, triplets of small gutta-percha markers were glued onto the soft tissues overlying the maxillary bone on the top and on both sides of the alveolar ridge to define a set of reproducible linear measurements in 11 planes. Image measurements were performed by 2 observers. The gold standard was determined by means of physical measurements with a caliper by 3 observers. RESULTS: The accuracy of the linear measurements was 0.35 +/- 1.31 mm (U70u) and 0.06 +/- 1.23 mm (U90u) forthe Somatom VolumeZoom, 0.24 +/- 1.20 mm (H60s) and 0.54 +/- 1.14 mm (H30s) for the Sensation 16, and -0.09 +/- 1.64 mm for the Accuitomo 3D. Statistical analysis with 2-way analysis of variance showed no significant inter- or intraobserver disagreement for the physical or the radiologic measurements. There was also no significant difference for the measurements on the different reconstruction filters. CONCLUSION: Both CBCT and MSCT yield submillimeter accuracy for linear measurements on an ex vivo specimen.
机译:目的:比较锥形束计算机断层扫描(CBCT)和多层CT(MSCT)进行线性颌骨测量的准确性。研究设计:用CBCT(Accuitomo 3D;日本京都盛田市)和MSCT(4层Somatom VolumeZoom和16层Somatom Sensation 16;西门子,德国埃尔兰根)对离体福尔马林固定的人上颌骨成像。使用不同的重建滤镜重建MSCT图像,以优化骨骼可视化(对于VolumeZoom为U70u和U90u,对于Sensation 16为H30s和H60s)。扫描前,将三连体小牙胶标记物粘贴到软组织上,该软组织覆盖在牙槽顶部和两侧的上颌骨上,以在11个平面上定义一组可重复的线性测量值。由2名观察者进行图像测量。金标准是由3位观察员使用卡尺通过物理测量来确定的。结果:Somatom VolumeZoom的线性测量精度为0.35 +/- 1.31 mm(U70u)和0.06 +/- 1.23 mm(U90u),0.24 +/- 1.20 mm(H60s)和0.54 +/- 1.14 mm(H30s) )适用于Sensation 16,而-0.09 +/- 1.64 mm适用于Accuitomo 3D。统计分析和方差的2-way分析表明,对于物理或放射学测量,观察者之间或观察者之间没有显着分歧。在不同的重建滤波器上进行的测量也没有显着差异。结论:CBCT和MSCT均可产生亚毫米级精度,可用于离体样本的线性测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号