首页> 外文期刊>Optics Letters >Common-path phase-shift interferometry surface plasmon resonance imaging system
【24h】

Common-path phase-shift interferometry surface plasmon resonance imaging system

机译:共路径相移干涉术表面等离子体共振成像系统

获取原文
获取原文并翻译 | 示例
           

摘要

Surface plasmon resonance (SPR) and common-path phase-shift interferometry (PSI) techniques are integrated in a biosensing imaging system for measuring the two-dimensional spatial phase variation caused by biomolecular interactions on a sensing chip without the need for additional labeling. The common-path PSI technique has the advantage of long-term stability, even when it is subjected to external disturbances. Hence the system meets the requirements of the real-time kinetic studies involved in biomolecular interaction analysis. The proposed SPR-PSI imaging system demonstrates a detection limit of a 2 x 10(-7) refractiveindex change, a long-term phase stability of 2.5 x 10(-4) pi rms for 4 h, and a spatial phase resolution of 10(-3) pi with a lateral resolution of 100 mu m. (c) 2005 Optical Society of America
机译:表面等离振子共振(SPR)和共通路径相移干涉术(PSI)技术集成在生物传感成像系统中,用于测量由传感芯片上生物分子相互作用引起的二维空间相变,而无需额外的标记。即使在受到外部干扰的情况下,公共路径PSI技术也具有长期稳定性的优点。因此,该系统满足参与生物分子相互作用分析的实时动力学研究的要求。拟议的SPR-PSI成像系统展示了2 x 10(-7)折射率变化的检测极限,2.5 x 10(-4)pi rms的长期相位稳定性4小时以及空间相位分辨率为10 (-3)圆周分辨率为100μm的pi。 (c)2005年美国眼镜学会