Repeatability of Retinal Macular Thickness Measurements in Patients with Clinically Significant Macular Edema Using Two Different Scanning Protocols of Spectralis Optical Coherence Tomography

机译:使用两种不同的Spectralis光学相干断层扫描技术对具有临床意义的黄斑水肿患者进行视网膜黄斑厚度测量的可重复性

获取原文
获取原文并翻译 | 示例
           

摘要

Objective: To determine the repeatability of Spectralis optical coherence tomography (OCT) retinal thickness measurements in diabetic patients with clinically significant macular edema (CSME) using two different scanning protocols. Methods: Seventy-one eyes of 71 diabetic patients with CSME were included in the study. Coefficients of repeatability and intrasession variation coefficients were tested with 20 × 15 degree raster scans consisting of 19 or 37 high-resolution line scans (15 or 8 frames per scan, respectively) that were repeated 2 times by 1 experienced examiner. The first scan was set as the reference scan; the second scan was the follow-up scan and was performed with the use of the follow-up mode. Results: The mean and standard deviation for the central foveal subfield (CSF) using the first scanning method was 404 ± 88 μm, while it was 399 ± 86 μm using the second protocol, which was not statistically significantly different (p = 0.35, paired test). Particularly examining the CSF, the coefficient of repeatability was 1.48 (6.00 μm) and 1.49 (5.95 μm) for the 19-and the 37-B-scan acquisition, respectively, showing a nonstatistically significant difference (p < 0.001). Conclusions: Retinal thickness measurements in diabetic patients with CSME are repeatable using both scanning protocols (19 or 37 B-scans) with Spectralis OCT. The repeatability of the retinal thickness measurement does not improve by increasing the number of B-scans from 19 to 37.
机译:目的:使用两种不同的扫描方案,确定具有临床显着黄斑水肿(CSME)的糖尿病患者的Spectralis光学相干断层扫描(OCT)视网膜厚度测量的可重复性。方法:纳入71例CSME糖尿病患者的71只眼。用20×15度光栅扫描测试重复性系数和会话期间变异系数,该扫描由19或37个高分辨率线扫描(每次扫描分别为15或8帧)组成,并由1位经验丰富的检查员重复2次。第一次扫描被设置为参考扫描。第二次扫描是跟进扫描,是使用跟进模式进行的。结果:使用第一种扫描方法的中央凹中心视野(CSF)的平均值和标准偏差为404±88μm,而使用第二种方案的平均值和标准差为399±86μm,在统计学上无显着差异(p = 0.35,配对测试)。特别是检查CSF,对于19扫描和37-B扫描采集,可重复性系数分别为1.48(6.00μm)和1.49(5.95μm),显示出无统计学意义的差异(p <0.001)。结论:使用Spectralis OCT的两种扫描方案(19或37 B扫描),都可以重复测量CSME糖尿病患者的视网膜厚度。通过将B扫描的数量从19增加到37不能提高视网膜厚度测量的可重复性。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号