Standards for radiation sterilization of health care products require evaluation of the potential for induced radioactivity for electron beam energies greater than 10MeV.Measured energy for a 10MeV accelerator may be slightly higher,which could fall under the requirement.The evaluation method in this paper uses a pathways analysis method to estimate individual dose,which is compared against the estimate of induced activity.Based on this comparison,the amount of induced activity at 10.4 MeV beam energy does not present a significant individual risk from exposure to the irradiated device.
机译:保健产品的辐射灭菌标准要求评估电子束能量大于10MeV的诱发放射性的潜力.10MeV加速器的测量能量可能略高,可能会低于要求。本文中的评估方法使用估计个体剂量的途径分析方法,将其与诱导活性的估计值进行比较。基于此比较,在10.4 MeV束能量下诱导活性的量不会因暴露于辐照装置而引起显着的个体风险。