首页> 外文期刊>Radiation Physics and Chemistry >Intensity variation study of the radiation field in a mammographic system using thermoluminescent dosimeters TLD-900 (CaSO4:Dy)
【24h】

Intensity variation study of the radiation field in a mammographic system using thermoluminescent dosimeters TLD-900 (CaSO4:Dy)

机译:使用热发光剂量计TLD-900(CaSO4:Dy)的乳腺摄影系统中辐射场的强度变化研究

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

This study presents the results of the intensity variation of the radiation field in a mammographic system using the thermoluminescent dosimeter TLD-900 (CaSO4:Dy). These TLDs were calibrated and characterized in an industrial X-ray system used for instruments calibration, in the energy range used in mammography. They were distributed in a matrix of 19 lines and five columns, covering an area of 18 cm x 8 cm in the center of the radiation field on the clinical equipment. The results showed a variation of the intensity probably explained by the non-uniformity of the field due to the heel effect.
机译:这项研究提出了使用热发光剂量计TLD-900(CaSO4:Dy)的乳腺摄影系统中辐射场强度变化的结果。这些TLD在乳房X线照相术使用的能量范围内,在用于仪器校准的工业X射线系统中进行了校准和表征。它们分布在19条线和5列的矩阵中,在临床设备的辐射场中心覆盖18 cm x 8 cm的区域。结果表明强度的变化可能是由于脚跟效应引起的电场不均匀所致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号