首页> 外文期刊>Radiation measurements >Development and intercomparison of a reusable radon-in-water standard
【24h】

Development and intercomparison of a reusable radon-in-water standard

机译:制定可重复使用的水中标准品并进行比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Radium-free standards are not readily available for proficiency testing of laboratories that conduct radon (Rn-222) analyses of water. For this study, 33 identical, reusable, radon-in-water standards were prepared using a Ra-226-loaded filter sandwiched in polyethylene sheeting. The Rn-222 concentrations in the Ra-226-free standards were measured by liquid scintillation counting and compared to 10 reference solutions containing Ra-226. The Rn-222 concentrations measured in the standards were consistent (standard deviation of <2%), but averaged substantially less than concentrations determined in the Ra-226 reference standards. At full ingrowth, 86% of the Rn-222 produced by the sandwiched Ra-226 sources emanated into the water. An intercomparison of radon-in-water standards, performed to examine the accuracy of analyses by commercial, government, and private companies, showed that 18 of the 21 participants reported concentrations within 25% of the known (693 Bq l(-1)).
机译:进行ra(Rn-222)水分析的实验室的能力测试尚无法获得不含镭的标准品。在这项研究中,使用夹在聚乙烯薄膜中的装有Ra-226的过滤器制备了33种相同的,可重复使用的水中in标准液。通过液体闪烁计数测量不含Ra-226的标准液中Rn-222的浓度,并将其与10种含Ra-226的参比溶液进行比较。标准中测得的Rn-222浓度是一致的(标准偏差<2%),但平均水平明显小于Ra-226参考标准中测定的浓度。完全向内生长时,夹在中间的Ra-226来源产生的Rn-222的86%散发到水中。对水ra含量标准进行的比较,以检查商业,政府和私人公司进行的分析的准确性,结果表明,在21名参与者中,有18名报告的浓度在已知值的25%以内(693 Bq l(-1)) 。