首页> 外文期刊>Radiotherapy and oncology: Journal of the European Society for Therapeutic Radiology and Oncology >A method to estimate the transit dose on the beam axis for verification of dose delivery with portal images.
【24h】

A method to estimate the transit dose on the beam axis for verification of dose delivery with portal images.

机译:一种估算射束轴上的通过剂量的方法,以验证带有门图像的剂量传递。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE AND BACKGROUND: A feasibility study is performed to evaluate the possibility of using the transit dose of portal images on the beam axis to measure the accuracy in dose delivery. The algorithm and the method are tested on a breast phantom and on patients with a breast disease. MATERIALS AND METHODS: To estimate the transit dose at various air gaps behind the patient, a method is proposed which applies, for a given air gap, the inverse square law to the primary component of the exit dose and an experimentally determined function for the scatter component of the exit dose. It is assumed that the primary component and the scattered component of the exit dose are given by the treatment planning system. The experimental function for the variation of the scattered component with the air gap, determined by phantom measurements, is modelled by an analytical function which contains only field size, air gap and one energy-dependent parameter. RESULTS: The measurements on the breast phantom yield a maximum deviation between measured and estimated transit doses of 4.5%. The mean deviation is 0.9% with a standard deviation of the distribution of 2.3%. In vivo diode measurements on the same phantom yield a maximum deviation of 2.7%. Transit dose measurements on the beam axis for 45 portal images of breast patients show a mean deviation of 0.0% between the measured transit dose and the estimated transit dose. The standard deviation of the distribution is 4.4%. The method seems to be very sensitive to patient positioning and to discrepancies in breast thicknesses used for treatment planning. CONCLUSION: Preliminary results on breast patients show that the method proposed to evaluate transit doses on the beam axis from portal images may be a valuable alternative to conventional in vivo exit dosimetry. The method can be implemented in a simple way and does not require additional time during the irradiation session, as exit dosimetry with diodes does. The transit dose is only considered in one point. Nevertheless, in the framework of quality assurance of treatment delivery, this study is an example of the possibilities of monitoring at the same time the visual evaluation of the irradiated volume as well as the dosimetric control (i.e. in Gy) of treatment delivery with portal images.
机译:目的和背景:进行了可行性研究,以评估使用射束轴上的门户图像的转运剂量来测量剂量输送的准确性的可能性。该算法和方法在乳腺模型和患有乳腺疾病的患者上进行了测试。材料与方法:为了估计患者背后各种气隙处的转运剂量,提出了一种方法,该方法对于给定的气隙,将平方反比定律应用于出射剂量的主要成分,并通过实验确定散射函数出口剂量的成分。假定出口剂量的主要成分和分散成分由治疗计划系统给出。由幻像测量确定的用于散射分量随气隙变化的实验函数是通过分析函数建模的,该函数仅包含场大小,气隙和一个与能量有关的参数。结果:对乳房幻象的测量在所测量的和估计的通过剂量之间产生最大偏差,为4.5%。平均偏差为0.9%,分布的标准偏差为2.3%。在同一体模上进行的体内二极管测量产生的最大偏差为2.7%。乳房患者的45张门静脉图像在光轴上的转运剂量测量结果显示,所测得的转运剂量和估计的转运剂量之间的平均偏差为0.0%。分布的标准偏差为4.4%。该方法似乎对患者的位置以及用于治疗计划的乳房厚度差异非常敏感。结论:对乳腺患者的初步结果表明,所提出的评估门图像在光轴上的转运剂量的方法可能是常规体内出口剂量测定的一种有价值的替代方法。该方法可以以简单的方式实现,并且在辐照过程中不需要额外的时间,就像二极管的出口剂量法一样。仅在一个点考虑过境剂量。然而,在治疗提供质量保证的框架内,本研究是通过门户图像同时监测照射量的目视评估以及治疗提供的剂量控制(即,以Gy计)的可能性的一个例子。 。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号