首页> 外文期刊>Radiography >Image quality of digital mammography images produced using wet and dry laser imaging systems
【24h】

Image quality of digital mammography images produced using wet and dry laser imaging systems

机译:使用湿式和干式激光成像系统产生的数字乳房X线照片的图像质量

获取原文
获取原文并翻译 | 示例
           

摘要

Abstract Introduction A study was carried out to compare the quality of digital mammographic images printed or processed by a wet laser imaging system and a dedicated mammographic dry laser imaging system.Matejjal and methods Digital images of a tissue equivalent breast phantom were obtained using a GE Senographe 2000D digital mammography system and different target/filter combinations of the X-ray tube. These images were printed on films using the Fuji FL-IM D wet laser imaging system and the Kodak DryView 8600 dry laser imaging system. The quality of images was assessed in terms of detectability of microcalcifications and simulated tumour masses by five radiologists. In addition, the contrast index and speed index of the two systems were measured using the step wedge in the phantom. The unpaired, unequal variance t-test was used to test any statistically significant differences.Results There were no significant (p < 0.05) differences between the images printed using the two systems in terms of microcalcification and tumour mass detectability. The wet system resulted in slightly higher contrast index while the dry system showed significantly higher speed index.Conclusion Both wet and dry laser imaging systems can produce mammography images of good quality on which 0.2 mm microcalcifications and 2 mm tumour masses can be detected. Dry systems are preferable due to the absence of wet chemical processing and solid or liquid chemical waste. The wet laser imaging systems, however, still represent a useful alternative to dry laser imaging systems for mammography studies.
机译:摘要简介进行了一项研究,比较湿式激光成像系统和专用的乳腺摄影干式激光成像系统打印或处理的数字化乳房X线照片的质量。 2000D数字乳腺摄影系统和X射线管的不同目标/滤镜组合。这些图像使用Fuji FL-IM D湿式激光成像系统和Kodak DryView 8600干式激光成像系统打印在胶片上。由五位放射科医生根据微钙化的可检测性和模拟的肿瘤块评估了图像的质量。另外,两个模型的对比度指数和速度指数是使用幻影中的阶梯楔形测量的。结果采用两种系统打印的图像之间在微钙化和肿瘤质量可检测性方面均无显着性差异(p <0.05)。湿式系统的对比度指数略高,而干式系统的速度指数明显更高。结论湿式和干式激光成像系统均可产生质量良好的乳房X线照片,可以检测到0.2毫米的微钙化和2毫米的肿瘤块。由于没有湿化学处理和固体或液体化学废料,因此优选干式系统。然而,湿式激光成像系统仍然是干式X射线成像系统在乳腺摄影研究中的有用替代品。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号