首页> 外文期刊>Rapid Communications in Mass Spectrometry: RCM >Transient signal isotope analysis: validation of the method for isotope signal synchronization with the determination of amplifier first-order time constants
【24h】

Transient signal isotope analysis: validation of the method for isotope signal synchronization with the determination of amplifier first-order time constants

机译:瞬态信号同位素分析:验证同位素信号同步方法并确定放大器的一阶时间常数

获取原文
获取原文并翻译 | 示例
           

摘要

RATIONALE: During transient signal acquisition by Multi-Collection Inductively Coupled Plasma Mass Spectrometry (MC-ICPMS), an isotope ratio increase or decrease (isotopic drift hereafter) is often observed which is related to the different time responses of the amplifiers involved in multi-collection. This isotopic drift affects the quality of the isotopic data and, in a recent study, a method of internal amplifier signal synchronization for isotope drift correction was proposed. In this work the determination of the amplifier time constants was investigated in order to validate the method of internal amplifier signal synchronization for isotope ratio drift correction.
机译:理由:在通过多集合电感耦合等离子体质谱仪(MC-ICPMS)采集瞬态信号期间,经常观察到同位素比率的增加或减少(此后为同位素漂移),这与参与多光谱分析的放大器的不同时间响应有关。采集。这种同位素漂移会影响同位素数据的质量,最近的研究提出了一种用于同位素漂移校正的内部放大器信号同步方法。在这项工作中,研究了确定放大器时间常数的方法,以验证用于同位素比漂移校正的内部放大器信号同步方法。