首页> 外文期刊>Radiology >Digital chest radiography with a solid-state flat-panel x-ray detector: contrast-detail evaluation with processed images printed on film hard copy.
【24h】

Digital chest radiography with a solid-state flat-panel x-ray detector: contrast-detail evaluation with processed images printed on film hard copy.

机译:带有固态平板X射线检测器的数字化胸部X射线照相:对比度细节评估,将处理后的图像打印在胶片硬拷贝上。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: To evaluate and compare human observer performance in a contrast-detail test by using postprocessed hard-copy images from a digital chest radiography system and conventional screen-film radiographs. MATERIALS AND METHODS: The digital radiography system is based on a large-area flat-panel x-ray detector with a structured cesium iodide scintillator layer and an amorphous silicon thin-film transistor array for image readout. Images of a contrast-detail phantom were acquired at two exposure levels by using two standard thoracic screen-film systems and the digital system at matched dose. By using images of the phantom processed with standard chest image postprocessing techniques, a four-alternative forced-choice observer perception study was performed, and the number of detectable test signals (disk-shaped objects 0.3-4.0 mm in diameter) was determined for each image type. RESULTS: On average, observers detected more test signals on digital images than on screen-film radiographs at all diameters up to 2.0 mm and an equivalent number at larger diameters. Test signals with lower inherent subject contrast were detected more readily on digital images than on screen-film images, even when x-ray exposure levels for the digital system were reduced by 20%. CONCLUSION: Observer performance in a contrast-detail detection task can be improved by using images acquired with the flat-panel digital chest radiography system as compared with those acquired with state-of-the-art screen-film combinations.
机译:目的:通过使用数字胸部放射线照相系统的后处理硬拷贝图像和常规的屏幕胶片放射线照相,在对比细节测试中评估和比较人类观察者的表现。材料与方法:数字射线照相系统基于具有结构化碘化铯闪烁体层和用于图像读取的非晶硅薄膜晶体管阵列的大面积平板X射线探测器。通过使用两个标准的胸部屏幕胶片系统和数字系统以匹配的剂量,在两个曝光水平下获得了对比度细节体模的图像。通过使用通过标准胸部图像后处理技术处理过的幻影图像,进行了四项强制选择观察者感知研究,并确定了每种信号的可检测测试信号(直径为0.3-4.0 mm的盘状物体)的数量。图像类型。结果:平均而言,在最大直径为2.0 mm的所有直径上,观察者在数字图像上比在屏幕胶片X射线照片上检测到的测试信号更多,而在较大直径下,则为同等数量。即使数字系统的X射线曝光水平降低了20%,在数字图像上也比在屏幕胶片图像上更容易检测到具有较低固有主体对比度的测试信号。结论:与使用最新的屏幕胶片组合获取的图像相比,使用平板数字胸部放射线照相系统获取的图像可以提高对比度细节检测任务中的观察者性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号