Sequence design and evaluation of the reproducibility of water-selective diffusionweighted imaging of the breast at 3 T

机译:3 T时乳房水选择性扩散加权成像的可重复性的序列设计和评估

获取原文
获取原文并翻译 | 示例
           

摘要

Diffusion measurements derived from breast MRI can be adversely affected by unwanted signals from abundant fatty tissues if they are not suppressed adequately. To minimize this undesired contribution, we designed and optimized a water-selective diffusion-weighted imaging (DWI) sequence, which relies on spectrally selective excitation on the water resonance, obviating the need for fat suppression. As this method is more complex than standard DWI methods,we also report a test–retest study to evaluate its reproducibility. In this study, a spectrally selective Gaussian pulse on water resonance was combined with a pair of slice-selective adiabatic refocusing pulses for water-only DWI. Field map-based shimming and manual determination of the center frequency were used for water selection. The selectivity of the excitation pulse was optimized by a spectrally selective spectroscopy sequence based on the same principles. A test–retest study of 10 volunteers in two separate visits was used to evaluate its reproducibility. Our results from all subjects showed high-quality diffusion-weighted images of the breast without fat contamination. Mean apparent diffusion coefficients for b = 0, 600 s/mm~2 and b = 50, 600 s/mm~2 all showed good reproducibility, as 95% confidence intervals of the apparent diffusion coefficients were 4 × 10~(–5)mm~2/s and 5 × 10–5mm~2/s and repeatability values were 1.09× 10~(–4) and 1.31× 10~(–4), respectively. In conclusion, water-selective DWI is a feasible alternative to standard methods of DWI based on fat suppression. The added complexity of the method does not compromise the reproducibility of diffusion measurements in the breast.
机译:如果无法充分抑制脂肪组织的有害信号,则源自乳房MRI的扩散测量结果可能会受到不利影响。为了最大程度地减少这种不希望有的贡献,我们设计并优化了水选择性扩散加权成像(DWI)序列,该序列依赖于水共振的光谱选择性激发,从而消除了对脂肪抑制的需求。由于此方法比标准DWI方法更复杂,因此我们还报告了一项重新测试研究,以评估其可重复性。在这项研究中,针对水共振的光谱选择性高斯脉冲与一对仅用于水的DWI的切片选择性绝热重聚焦脉冲相结合。基于场图的匀场和手动确定中心频率用于选水。通过基于相同原理的光谱选择性光谱序列优化了激发脉冲的选择性。在两次单独的访问中对10名志愿者进行了一次测试-再测试研究,以评估其可重复性。我们所有受试者的结果均显示了高质量的乳房扩散扩散加权图像,没有脂肪污染。 b = 0,600 s / mm〜2和b = 50,600 s / mm〜2时的平均表观扩散系数均具有良好的重现性,因为表观扩散系数的95%置信区间为4×10〜(–5) mm〜2 / s和5×10-5mm〜2 / s,重复性值分别为1.09×10〜(–4)和1.31×10〜(–4)。总之,水选择性DWI是基于脂肪抑制的DWI标准方法的可行替代方法。该方法增加的复杂性不会影响乳房中扩散测量的可重复性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号