Understanding and Mitigating Unexpected Artifacts in Head CTs: A Practical Experience

机译:了解和缓解头部CT的意外伪像:实践经验

获取原文
获取原文并翻译 | 示例
       

摘要

CT artifacts may seriously degrade image quality by obscuring or mimicking pathology, and thus be the source of erroneous diagnosis. It is imperative for the neuroradiologist to understand why artifacts occur and how they can be prevented or suppressed. This review presents a potpourri of common as well as rarely mentioned artifacts borne out of clinical experience in a multifaceted academic neuroradiology practice. The root causes of these artifacts are classified into scanner limitation, technologist practice, and protocol implementation. The discussion of each artifact unfolds in a 3-step format: clinical setting, understanding, and mitigation. Considerations of both proper technique and technique failure within a pragmatic framework of applied clinical practice are essential for recognition and avoidance of such artifacts.
机译:CT伪影可能会掩盖或模仿病理现象,从而严重降低图像质量,从而成为错误诊断的根源。神经放射科医生必须了解为什么会出现伪影以及如何预防或抑制伪影。这篇综述介绍了从多方面学术神经放射学实践中的临床经验中发现的常见的和很少提及的文物。这些伪像的根本原因可分为扫描仪限制,技术人员实践和协议实施。对每个工件的讨论分三步进行:临床设置,理解和缓解。在实用的临床实践实践框架内考虑适当的技术和技术失败对于识别和避免此类伪影至关重要。