首页> 外文期刊>MTT: Manufacturing Technology Today >CHARGE AMPLIFIERS CALIBRATION IN VIBRATION METROLOGY AT NPLI
【24h】

CHARGE AMPLIFIERS CALIBRATION IN VIBRATION METROLOGY AT NPLI

机译:NPLI振动计量中的电荷放大器校准

获取原文
获取原文并翻译 | 示例
           

摘要

The present work discusses the methodology and procedure followed at NPLI for calibration of charge amplifiers. The charge amplifiers are calibrated conventionally using a 1 nF standard capacitor and sine generator for gain measurements. An automated calibration system using B&K Calibration software in conjunction with a calibrated 1 nF precision adaptor controlled by Pulse Analyzer is also used for precision calibration of charge amplifiers for gain and phase measurements from 0.5 Hz to 20 kHz. The uncertainty achieved in gain calibration is reported in the present work.
机译:本工作讨论了在NPLI进行电荷放大器校准的方法和步骤。电荷放大器通常使用1 nF标准电容器和正弦发生器进行校准,以进行增益测量。结合使用B&K校准软件和由Pulse Analyzer控制的经过校准的1 nF精密适配器的自动校准系统,还可用于电荷放大器的精确校准,以进行0.5 Hz至20 kHz的增益和相位测量。在本工作中报告了增益校准中实现的不确定性。