首页> 外文期刊>Medical Physics >Simulation study of a quasi-monochromatic beam for x-ray computed mammotomography.
【24h】

Simulation study of a quasi-monochromatic beam for x-ray computed mammotomography.

机译:用于X射线计算机乳腺摄影的准单色光束的仿真研究。

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this simulation study was to evaluate the feasibility, benefits, and potential operating parameters of a quasi-monochromatic beam from a tungsten-target x-ray source yielding projection images. The application is intended for newly developed cone beam computed mammotomography (CmT) of an uncompressed breast. The value of a near monochromatic x-ray source for a fully 3D CmT application is the expected improved ability to separate tissues with very small differences in attenuation coefficients. The quasi-monochromatic beam is expected to yield enhanced tomographic image quality along with a low dose, equal to or less than that of dual view x-ray mammography. X-ray spectra were generated with a validated projection x-ray simulation tool (XSpect) for a range of tungsten tube potentials (40-100 kVp), filter materials (Z=51-65), and filter thicknesses (10th to 1000th value layer determined at 60 kVp). The breast was modeled from ICRU-44 breast tissue specifications, and a breast lesion was modeled as a 0.5 cm thick mass. The detector was modeled as a digital flat-panel detector with a 0.06 cm thick CsI x-ray absorption layer. Computed figures of merit (FOMs) included the ratio of mean beam energy post-breast to pre-breast and the ratio of lesion contrasts for edge-located and center-located lesions as indices of breast beam hardening, and SNR2/exposure and SNR2/dose as indices of exposure and dose efficiencies. The impact of optimization of these FOMs on lesion contrast is also examined. For all simulated filter materials at each given attenuation thickness [10th, 100th, 500th, 1000th value layers (VLs)], the mean and standard deviation of the pre-breast spectral full-width at tenth-maximum (FWTM) were 16.1 +/- 2.4, 10.3 +/- 2.2, 7.3 +/- 1.4, and 6.5 +/- 1.5 keV, respectively. The change in beam width at the tenth maximum from pre-breast to post-breast spectra ranged from 4.7 to 1.1 keV, for the thinnest and thickest filters, respectively. The higher Z filters (Z=57-63) produced a quasi-monochromatic beam that allowed the widest tube potential operating range (50-70 kVp) while maintaining minimal beam hardening and maximal SNR2/exposure and SNR2/dose, and providing a contrast greater than that obtained in the unfiltered case. Figures of merit improved with increasing filter thickness, with diminishing returns beyond the 500th value layer attenuation level. Operating parameters required to produce optimal spectra, while keeping exposures equal to that of dual view mammography, are within the capability of the commercial x-ray tube proposed for our experimental study, indicating that use of these highly attenuating filters is viable. Additional simulations comparing Mo/Mo, Mo/Rh, and W/Rh target/filter combinations indicate that they exhibit significantly lower SNR2/exposure than the present approach, precluding them from being used for computed mammotomography, while maintaining dose limitations and obtaining sufficient SNR. Beam hardening was also much higher in the existing techniques(17%-42%) than for our technique (2%). Simulations demonstrate that this quasi-monochromatic x-ray technique may enhance tissue separation for a newly developed cone beam computed mammotomography application for an uncompressed breast.
机译:本模拟研究的目的是评估来自钨靶x射线源产生投影图像的准单色光束的可行性,优势和潜在的工作参数。该应用程序旨在用于未受压乳房的最新开发的锥形束计算机X线乳腺摄影(CmT)。对于完全3D CmT应用而言,近乎单色的X射线源的价值在于,可以预期在分离衰减系数非常小的差异的组织方面提高分离能力。准单色光束有望以较低的剂量产生增强的X射线断层扫描图像质量,该剂量等于或小于双透视X线乳房X线照相术的剂量。 X射线光谱是使用经过验证的投影X射线模拟工具(XSpect)生成的,用于一系列钨管电势(40-100 kVp),过滤材料(Z = 51-65)和过滤厚度(第10至1000值)层在60 kVp下确定)。根据ICRU-44乳房组织规格对乳房进行建模,并将乳房病变模型化为0.5厘米厚的肿块。该检测器被建模为具有0.06 cm厚的CsI x射线吸收层的数字平板检测器。计算的品质因数(FOM)包括乳房后与乳房前的平均束能量之比以及位于边缘和中心位置的病变的病变对比度之比作为乳腺束硬化的指标,以及SNR2 /暴露和SNR2 /剂量作为暴露量和剂量效率的指标。还检查了这些FOM的优化对病变对比度的影响。对于每个给定衰减厚度[第10、100、500、1000值层(VLs)]的所有模拟滤波器材料,在最大最大值(FWTM)处的前乳房光谱全宽的平均值和标准偏差为16.1 + / -分别为2.4、10.3 +/- 2.2、7.3 +/- 1.4和6.5 +/- 1.5 keV。对于最薄和最厚的滤光片,从乳房前到乳房后光谱的第十个最大值的束宽变化范围为4.7到1.1 keV。较高的Z滤波器(Z = 57-63)产生准单色光束,该光束允许最宽的管电位工作范围(50-70 kVp),同时保持最小的光束硬化和最大的SNR2 /曝光量和SNR2 /剂量,并提供对比度大于未过滤情况下获得的结果。品质因数随滤波器厚度的增加而提高,返回值减小到第500值层衰减水平以上。产生最佳光谱所需的操作参数,同时保持与双乳腺X线摄影相等的曝光量,是在我们的实验研究中建议的商用X射线管的能力之内,表明使用这些高衰减滤光片是可行的。比较Mo / Mo,Mo / Rh和W / Rh目标/滤光片组合的其他模拟表明,与目前的方法相比,它们的SNR2 /曝光量显着降低,从而阻止了它们被用于计算机X线体层摄影术,同时保持了剂量限制并获得了足够的SNR。 。在现有技术中,束硬化也要比我们的技术(2%)高得多(17%-42%)。仿真表明,对于未经压缩的乳房,这种准单色X射线技术可为新开发的锥形束计算机X线断层摄影术应用增强组织分离。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号