摘要

We have designed and constructed a wearable device for skin-contact thermography [Giansanti D, Maccioni G. Development and testing of a wearable integrated thermometer sensor for skin-contact thermography. Med Eng Phys 2007;29(5):556-65] that uses 16 integrated silicon sensors with a thermal resolution of 0.03 degrees C and a spatial resolution of 1.6 x 10(-5)m(2). In order to perform a dynamic bench test, a thermographic phantom was designed and constructed that permits the imposition of 16 independent thermal waveforms to the device. The phantom was based on 16 small black bodies units designed with a 10(-4) m thin layer of black conductive gum. An infrared system was used in closed loop to correct in real time the error between the theoretical and the actual temperature imposed. The methodology proved to be adequate for bench testing the device. The difference between the imposed and the theoretical waveforms was never greater than 0.85 x 10(-2) degrees C in mean value, as measured with the infrared(IR) system.
机译:我们已经设计并构建了一种用于皮肤接触热敏成像的可穿戴设备[Giansanti D,Maccioni G.开发和测试了一种用于皮肤接触热敏成像的可穿戴一体式温度计传感器。 Med Eng Phys 2007; 29(5):556-65],它使用16个集成的硅传感器,其热分辨率为0.03摄氏度,空间分辨率为1.6 x 10(-5)m(2)。为了执行动态工作台测试,设计并构造了一个热成像体模,该体模允许向设备施加16个独立的热波形。幻影基于16个黑色小物体,设计有10(-4)m薄的黑色导电胶层。在闭环中使用红外系统实时校正理论温度和实际温度之间的误差。实践证明该方法足以对设备进行基准测试。用红外(IR)系统测得,所施加波形与理论波形之间的差异均不大于0.85 x 10(-2)摄氏度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号