首页> 外文期刊>Medical Physics >Calibration of EBT2 film using a red-channel PDD method in combination with a modified three-channel technique
【24h】

Calibration of EBT2 film using a red-channel PDD method in combination with a modified three-channel technique

机译:使用红通道PDD方法结合改进的三通道技术对EBT2胶片进行校准

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: Ashland Inc. EBT2 and EBT3 films are widely used in quality assurance for radiation therapy; however, there remains a relatively high degree of uncertainty [B. Hartmann, M. Martisikova, and O. Jakel, "Homogeneity of Gafchromic EBT2 film," Med. Phys. 37, 1753-1756 (2010)]. Micke et al. (2011) recently improved the spatial homogeneity using all color channels of a flatbed scanner; however, van Hoof et al. (2012) pointed out that the corrected nonuniformity still requires further investigation for larger fields. To reduce the calibration errors and the uncertainty, the authors propose a new red-channel percentage-depth-dose method in combination with a modified three-channel technique.
机译:目的:Ashland Inc. EBT2和EBT3膜广泛用于放射治疗的质量保证;但是,仍然存在相对较高的不确定性[B。 Hartmann,M。Martisikova和O. Jakel,“全色EBT2膜的同质性”,医学。物理37,1753-1756(2010)]。 Micke等。 (2011年)最近使用平板扫描仪的所有颜色通道改善了空间均匀性;但是,van Hoof等人。 (2012)指出,校正后的非均匀性仍然需要对更大的领域进行进一步研究。为了减少校准误差和不确定性,作者提出了一种新的红色通道百分比深度剂量方法,并结合了改进的三通道技术。