A statistical approach to infer the minimum setup distance of a well chamber to the wall or to the floor for 192Ir HDR calibration.

机译:一种统计方法,可推断出用于192Ir HDR校准的井腔室到墙壁或地板的最小安装距离。

获取原文
获取原文并翻译 | 示例
           

摘要

We propose a new method based on statistical analysis technique to determine the minimum setup distance of a well chamber used in the calibration of 192Ir high dose rate (HDR). The chamber should be placed at least this distance away from any wall or from the floor in order to mitigate the effect of scatter. Three different chambers were included in this study, namely, Sun Nuclear Corporation, Nucletron, and Standard Imaging. The results from this study indicated that the minimum setup distance varies depending on the particular chamber and the room architecture in which the chamber was used. Our result differs from that of a previous study by Podgorsak et al. [Med. Phys. 19, 1311-1314 (1992)], in which 25 cm was suggested, and also differs from that of the International Atomic Energy Agency (IAEA)-TECDOC-1079 report, which suggested 30 cm. The new method proposed in this study may be considered as an alternative approach to determine the minimum setup distance of a well-type chamber used in the calibration of 192Ir HDR.
机译:我们提出了一种基于统计分析技术的新方法,以确定用于校准192Ir高剂量率(HDR)的井室的最小安装距离。腔室应放置在距任何墙壁或地板至少此距离的位置,以减轻散射的影响。这项研究包括三个不同的室,分别是Sun Nuclear Corporation,Nucletron和Standard Imaging。这项研究的结果表明,最小安装距离取决于特定的腔室和使用该腔室的房间结构。我们的结果与Podgorsak等人先前的研究不同。 [医生。物理19,1311-1314(1992)],其中建议使用25厘米,也不同于国际原子能机构(IAEA)-TECDOC-1079报告的建议使用30厘米。这项研究中提出的新方法可以被视为确定192Ir HDR校准中使用的井式腔室最小设置距离的替代方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号