Rapid radiographic film calibration for IMRT verification using automated MLC fields.

机译:使用自动MLC字段对IMRT进行快速射线照相胶片校准。

获取原文
获取原文并翻译 | 示例

摘要

A method for measuring a film sensitometric curve using a single sheet of film exposed with a two field step-and-shoot MLC treatment was developed and tested with Kodak XV2 and EDR2 films. With this technique a film sensitometric curve can be completed in only 10 minutes, making it practical to generate new film calibrations daily. This method is applicable to film calibrations for all purposes, but is particularly useful in IMRT treatment verification due to the method's use of small fields. This method agrees with the traditional large-field multifilm calibration within 0.5% and will produce sensitometric curves with errors less than 1% throughout the dose range, including uncertainties in dose delivery, film response, and optical density measurements. OD values for XV2 and EDR2 films were consistent in the middle of exposure areas at high depths, but the XV2 film penumbra regions showed large amounts of over-response as the calibration depth increased. If XV2 film is used for IMRT treatment verification, it is necessary to reduce the fluence of low energy photons in areas around the film by using thin lead shields. EDR2 film was shown to have minimal energy dependence, as it accurately represented penumbra areas and yielded identical sensitometric curves generated with 6 and 18 MV photons. However, its darker tint may make it more sensitive to scanning laser film digitizers' horizontal nonuniformities. This single film method proved to be superior to the traditional calibration method and allows fast daily calibrations of films for highly accurate IMRT delivery verifications.
机译:柯达XV2和EDR2胶片的开发和测试方法是,使用单张胶片进行两场分步傻瓜式MLC处理曝光后,测量胶片感光度曲线。使用此技术,只需10分钟即可完成胶片感光度曲线,因此每天进行新的胶片校准非常实用。该方法适用于所有目的的胶片校准,但由于该方法使用的是小视野,因此在IMRT处理验证中特别有用。该方法与传统的大视野多层膜校准相吻合,误差在0.5%以内,在整个剂量范围内,其灵敏度曲线均误差小于1%,包括剂量输送,膜响应和光密度测量的不确定性。 XV2和EDR2胶片的OD值在高深度的曝光区域的中部是一致的,但是XV2胶片半影区域随着校准深度的增加而显示出大量的过度响应。如果将XV2膜用于IMRT处理验证,则必须使用薄的铅屏蔽层来减少低能量光子在膜周围区域的能量通量。 EDR2膜被证明具有最小的能量依赖性,因为它可以精确地表示半影区并产生由6和18 MV光子产生的相同的感光度曲线。但是,其较深的色彩可能使其对扫描激光胶片数字化仪的水平不均匀性更加敏感。事实证明,这种单胶片方法优于传统的校准方法,并且可以对胶片进行快速的每日校准,以进行高精度的IMRT交付验证。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号