Accuracy estimation for projection-to-volume targeting during rotational therapy: a feasibility study.

机译:旋转疗法中以投射为目标的准确度估算:一项可行性研究。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: Estimating motion and deformation parameters from a series of projection radiographs acquired during arc therapy using a reference CT volume has become a promising technique for targeting treatment. The purpose of this work is to investigate the influence of rotational arc length on maximum achievable accuracy of motion estimation. METHODS: The projection-to-volume alignment procedure used a nonrigid model to describe motion in thorax area, a cost function consisting of a least-squared error metric and a simple regularizer that encourages local invertibility, and a four-level multiresolution scheme with a conjugate gradient method to optimize the cost function. The authors tested both small and large scale deformations typically found in the thorax of a radiotherapy patient at different breathing states and limited-angle scans of six angular widths (12 degrees, 18 degrees, 24 degrees, 36 degrees, 60 degrees, and 90 degrees) centered at three angles (0 degrees, 45 degrees, and 90 degrees). RESULTS: The experiments illustrate the potential accuracy of limited-angle projection-to-volume alignment. Registration accuracy can be sensitive to angular center, tends to be lower along direction of the projection set, and tends to decrease away from the rotation center. The studies of small as well as large but realistically scaled deformations show similar dependencies. These trends appear to have fairly low sensitivity to quantum noise effects. CONCLUSIONS: There is potentially sufficient information present in a small spread of projections to monitor the configuration of reasonably high contrast tumors without implanted markers.
机译:目的:从使用参考CT容积在弧光治疗期间获得的一系列投影射线照片估计运动和变形参数已成为一种有针对性的靶向治疗技术。这项工作的目的是研究旋转弧长对运动估计最大可达到精度的影响。方法:体积投影投影法使用非刚性模型描述胸部区域的运动,成本函数包括最小二乘误差度量和鼓励局部可逆性的简单正则化函数,以及四级多分辨率方案,其中包括共轭梯度法优化成本函数。作者测试了在不同呼吸状态和六个角度宽度(12度,18度,24度,36度,60度和90度)的有限角度扫描中通常在放射治疗患者的胸部中发现的小尺寸和大尺寸变形)以三个角度(0度,45度和90度)为中心。结果:实验说明了有限角度投影到体积对齐的潜在准确性。套准精度可能对角中心敏感,沿投影集的方向趋于降低,并且趋于远离旋转中心而降低。对大小形变但实际缩放的变形的研究也显示出相似的依赖性。这些趋势似乎对量子噪声效应的灵敏度相当低。结论:在一小部分投影中可能存在足够的信息,以监测没有植入标记物的合理高对比度肿瘤的形态。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号