For commissioning a linear accelerator for clinical use, medical physicists are faced with many challenges including the need for precision, a variety of testing methods, data validation, the lack of standards, and time constraints. Since commissioning beam data are treated as a reference and ultimately used by treatment planning systems, it is vitally important that the collected data are of the highest quality to avoid dosimetric and patient treatment errors that may subsequently lead to a poor radiation outcome. Beam data commissioning should be performed with appropriate knowledge and proper tools and should be independent of the person collecting the data. To achieve this goal, Task Group 106 (TG-106) of the Therapy Physics Committee of the American Association of Physicists in Medicine was formed to review the practical aspects as well as the physics of linear accelerator commissioning. The report provides guidelines and recommendations on the proper selection of phantoms and detectors, setting up of a phantom for data acquisition (both scanning and no-scanning data), procedures for acquiring specific photon and electron beam parameters and methods to reduce measurement errors (<1%), beam data processing and detector size convolution for accurate profiles. The TG-106 also provides a brief.discussion on the emerging trend in Monte Carlo simulation techniques in photon and electron beam commissioning. The procedures described in this report should assist a qualified medical physicist in either measuring a complete set of beam data, or in verifying a subset of data before initial use or for periodic quality assurance measurements. By combining practical experience with theoretical discussion, this document sets a new standard for beam data commissioning.
机译:为了调试用于临床的线性加速器,医学物理学家面临许多挑战,包括对精度的要求,各种测试方法,数据验证,缺乏标准和时间限制。由于调试光束数据将作为参考,并最终由治疗计划系统使用,因此至关重要的是,所收集的数据必须具有最高的质量,以避免剂量和患者治疗错误,这些错误可能随后导致不良的放射结果。光束数据的调试应使用适当的知识和适当的工具进行,并且应独立于收集数据的人员。为了实现这一目标,成立了美国医学物理学会治疗物理学委员会的任务组106(TG-106),以审查线性加速器调试的实际情况和物理学。该报告提供了有关正确选择体模和检测器,设置体模以进行数据采集(扫描和非扫描数据),获取特定光子和电子束参数的程序以及减少测量误差的方法的指南和建议(< 1%),光束数据处理和检测器尺寸卷积以获得精确轮廓。 TG-106还简要介绍了蒙特卡罗模拟技术在光子和电子束调试中的新兴趋势。本报告中所述的程序应协助合格的医学物理学家测量整套射束数据,或在初次使用之前或进行定期质量保证测量时验证数据的子集。通过将实践经验与理论讨论相结合,该文档为光束数据调试树立了新的标准。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号