...
首页> 外文期刊>Medical Physics >A phantom evaluation of a stereo-vision surface imaging system for radiotherapy patient setup.
【24h】

A phantom evaluation of a stereo-vision surface imaging system for radiotherapy patient setup.

机译:立体视觉表面成像系统用于放射治疗患者装置的幻像评估。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

External beam irradiation requires precise positioning of the target relative to the treatment planning coordinate system. A three-dimensional (3D) surface imaging system for patient positioning has recently been installed in one of our linear accelerator (linac) rooms. The device utilizes close-range photogrammetry to generate a 3D model of the patient's surface. This geometric model can be made to look like a digital camera image if wrapped with a gray-level image (texture mapping) that shows surface coloration. The system is calibrated to the linac coordinate system and has been designed as a patient setup device. To reproduce patient position in fractionated radiotherapy, the daily patient surface model is registered to a previously recorded reference surface. Using surface registration, the system calculates the rigid-body transformation that minimizes the distance between the treatment and the reference surface models in a region-of-interest (ROI). This transformation is expressed as a set of newcouch coordinates at which the patient position best matches with the reference data. If respiratory motion is a concern, the surface can be obtained with a gated acquisition at a specified phase of the respiratory cycle. To analyze the accuracy of the system, we performed several experiments with phantoms to assess stability, alignment accuracy, precision of the gating function, and surface topology. The reproducibility of surface measurements was tested for periods up to 57 h. Each recorded frame was registered to the reference surface to calculate the required couch adjustment. The system stability over this time period was better than 0.5 mm. To measure the accuracy of the system to detect and quantify patient shift relative to a reference image, we compared the shift detected by the surface imaging system with known couch transitions in a phantom study. The maximum standard deviation was 0.75 mm for the three translational degrees of freedom, and less than 0.1 degrees for each rotation. Surface model precision was tested against computed tomography (CT)-derived surface topology. The root-mean-square rms of the distance between the surfaces was 0.65 mm, excluding regions where beam hardening caused artifacts in the CT data. Measurements were made to test the gated acquisition mode. The time-dependent amplitude was measured with the surface imaging system and an established respiratory gating system based on infrared (IR)-marker detection. The measured motion trajectories from both systems were compared to the known trajectory of the stage. The standard deviations of the amplitude differences to the motor trajectory were 0.04 and 0.15 mm for the IR-marker system and the 3D surface imaging system, respectively. A limitation of the surface-imaging device is the frame rate of 6.5 Hz, because rapid changes of the motion trajectory cannot be detected. In conclusion, the system is accurate and sufficiently stable to be used in the clinic. The errors computed when comparing the surface model with CT geometry were submillimeter, and deviations in the alignment and gating-signal tests were of the same magnitude.
机译:外部束照射需要相对于治疗计划坐标系精确定位目标。最近在我们的线性加速器(直线加速器)室中安装​​了用于患者定位的三维(3D)表面成像系统。该设备利用近距离摄影测量法生成患者表面的3D模型。如果用显示表面着色的灰度级图像(纹理映射)包裹起来,可以使此几何模型看起来像数码相机图像。该系统已根据直线加速器坐标系进行了校准,并已被设计为患者设置设备。为了在分级放疗中重现患者的位置,将每日患者表面模型注册到先前记录的参考表面。该系统使用表面配准计算出一个刚体变换,该变换使在感兴趣区域(ROI)中处理与参考表面模型之间的距离最小。此转换表示为一组新沙发坐标,患者位置处的新沙发坐标与参考数据最匹配。如果需要考虑呼吸运动,则可以在呼吸周期的指定阶段通过门控采集获得表面。为了分析系统的精度,我们使用幻像进行了几次实验,以评估稳定性,对准精度,门控功能的精度和表面拓扑。测试了表面测量的可重复性长达57小时。将每个记录的帧配准到参考表面,以计算所需的卧榻调节。在此期间的系统稳定性优于0.5毫米。为了测量检测和量化患者相对于参考图像的偏移的系统的准确性,我们在幻像研究中将表面成像系统检测到的偏移与已知的卧榻转换进行了比较。三个平移自由度的最大标准偏差为0.75 mm,每次旋转小于0.1度。针对源自计算机断层扫描(CT)的表面拓扑测试了表面模型的精度。表面之间的距离的均方根均方根为0.65 mm,其中不包括光束硬化导致CT数据中出现伪影的区域。进行测量以测试门控采集模式。使用表面成像系统和基于红外(IR)标记检测的已建立的呼吸门控系统测量随时间变化的幅度。将两个系统中测得的运动轨迹与平台的已知轨迹进行比较。对于IR标记系统和3D表面成像系统,振幅差相对于电机轨迹的标准偏差分别为0.04和0.15 mm。表面成像设备的局限性是6.5 Hz的帧频,因为无法检测到运动轨迹的快速变化。总之,该系统准确且足够稳定,可用于临床。将表面模型与CT几何形状进行比较时计算出的误差为亚毫米级,并且对准和门控信号测试中的偏差幅度相同。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号