A phantom evaluation of Sentinel, a commercial laser/camera surface imaging system for patient setup verification in radiotherapy

机译:Sentinel的幻影评估,Sentinel是商用激光/相机表面成像系统,用于放射治疗中的患者设置验证

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Purpose: The aim of this study is to investigate the performance of Sentinel, a patient setup verification device based on a laser/camera system, when used on rigid-body phantoms. Methods: The Sentinel system consists of a scanner unit, containing the laser and the camera, and surface-matching registration software. For the registration procedure, both a computed tomography (CT) and a Sentinel image can be used as a reference. Tests were performed on phantoms using an Elekta Synergy beam modulator Linac equipped with a cone beam CT (CBCT), a HexaPOD RT couch top, and an iViewGT portal imaging system. Experiments conducted in this study tested reproducibility of the Sentinel surface acquisition and of set-up procedure, accuracy in quantifying known phantom mispositioning, and compared Sentinel, CBCT, and portal imaging system performance. Results: Reproducibility of surface acquisition and setup procedure was better than 0.5 mm and 0.5 and 1 mm and 0.4, respectively. The system accuracy was better than 1 mm and 1 when a Sentinel image was used as reference. A global worsening of Sentinel performance was observed using as reference an external surface extracted from CT study. This effect is probably due to small differences in considered surfaces, caused by different imaging modalities. The results obtained by testing the system on rigid phantoms were comparable to those obtained using CBCT and better than those obtained with conventional portal imaging systems. Conclusions: The Sentinel setup verification device is a reproducible and consistent system able to detect misalignments with accuracy better than 1 mm and 1. When tested on rigid body phantoms, Sentinel and CBCT performed similarly. When compared to portal imaging, both Sentinel and CBCT were more accurate.
机译:目的:本研究的目的是研究Sentinel(基于激光/相机系统的患者设置验证设备)在刚体模型上使用时的性能。方法:Sentinel系统由一个扫描仪单元(包含激光和照相机)和表面匹配注册软件组成。对于注册过程,计算机断层扫描(CT)和Sentinel图像均可用作参考。使用配备了锥形束CT(CBCT),HexaPOD RT沙发顶和iViewGT门禁成像系统的Elekta Synergy光束调制器Linac对人体模型进行测试。在这项研究中进行的实验测试了Sentinel表面采集和设置程序的可重复性,量化了已知幻影错位的准确性,并比较了Sentinel,CBCT和门禁成像系统的性能。结果:表面采集和安装程序的可重复性分别优于0.5 mm,0.5和1 mm和0.4。当使用Sentinel图像作为参考时,系统精度优于1 mm和1。使用从CT研究中提取的外部表面作为参考,可以观察到Sentinel性能的整体恶化。这种影响可能是由于不同的成像方式导致所考虑的表面存在细微差异。通过在刚性体模上测试系统而获得的结果与使用CBCT获得的结果相当,并且优于使用常规门禁成像系统获得的结果。结论:Sentinel设置验证设备是一种可重现且一致的系统,能够检测精度超过1 mm和1的未对准。在刚体体模上进行测试时,Sentinel和CBCT的性能类似。与门静脉成像相比,Sentinel和CBCT都更准确。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号