首页> 外文期刊>Medical Physics >CT head-scan dosimetry in an anthropomorphic phantom and associated measurement of ACR accreditation-phantom imaging metrics under clinically representative scan conditions
【24h】

CT head-scan dosimetry in an anthropomorphic phantom and associated measurement of ACR accreditation-phantom imaging metrics under clinically representative scan conditions

机译:在具有临床代表性的扫描条件下,拟人化体模中的CT头扫描剂量测定以及ACR认证体模成像指标的相关测量

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To measure radiation absorbed dose and its distribution in an anthropomorphic head phantom under clinically representative scan conditions in three widely used computed tomography (CT) scanners, and to relate those dose values to metrics such as high-contrast resolution, noise, and contrast-to-noise ratio (CNR) in the American College of Radiology CT accreditation phantom. Methods: By inserting optically stimulated luminescence dosimeters (OSLDs) in the head of an anthropomorphic phantom specially developed for CT dosimetry (University of Florida, Gainesville), we measured dose with three commonly used scanners (GE Discovery CT750 HD, Siemens Definition, Philips Brilliance 64) at two different clinical sites (Walter Reed National Military Medical Center, National Institutes of Health). The scanners were set to operate with the same data-acquisition and image-reconstruction protocols as used clinically for typical head scans, respective of the practices of each facility for each scanner. We also analyzed images of the ACR CT accreditation phantom with the corresponding protocols. While the Siemens Definition and the Philips Brilliance protocols utilized only conventional, filtered back-projection (FBP) image-reconstruction methods, the GE Discovery also employed its particular version of an adaptive statistical iterative reconstruction (ASIR) algorithm that can be blended in desired proportions with the FBP algorithm. We did an objective image-metrics analysis evaluating the modulation transfer function (MTF), noise power spectrum (NPS), and CNR for images reconstructed with FBP. For images reconstructed with ASIR, we only analyzed the CNR, since MTF and NPS results are expected to depend on the object for iterative reconstruction algorithms. Results: The OSLD measurements showed that the Siemens Definition and the Philips Brilliance scanners (located at two different clinical facilities) yield average absorbed doses in tissue of 42.6 and 43.1 mGy, respectively. The GE Discovery delivers about the same amount of dose (43.7 mGy) when run under similar operating and image-reconstruction conditions, i.e., without tube current modulation and ASIR. The image-metrics analysis likewise showed that the MTF, NPS, and CNR associated with the reconstructed images are mutually comparable when the three scanners are run with similar settings, and differences can be attributed to different edge-enhancement properties of the applied reconstruction filters. Moreover, when the GE scanner was operated with the facility's scanner settings for routine head exams, which apply 50% ASIR and use only approximately half of the 100%-FBP dose, the CNR of the images showed no significant change. Even though the CNR alone is not sufficient to characterize the image quality and justify any dose reduction claims, it can be useful as a constancy test metric. Conclusions: This work presents a straightforward method to connect direct measurements of CT dose with objective image metrics such as high-contrast resolution, noise, and CNR. It demonstrates that OSLD measurements in an anthropomorphic head phantom allow a realistic and locally precise estimation of magnitude and spatial distribution of dose in tissue delivered during a typical CT head scan. Additional objective analysis of the images of the ACR accreditation phantom can be used to relate the measured doses to high contrast resolution, noise, and CNR.
机译:目的:在三台广泛使用的计算机断层扫描(CT)扫描仪中,在临床代表性扫描条件下,测量拟人化人体模型中的辐射吸收剂量及其分布,并将这些剂量值与诸如高对比度分辨率,噪声和对比度的指标相关联美国放射学院CT认证模型中的信噪比(CNR)。方法:通过将光激发发光剂量计(OSLD)插入专门为CT剂量学开发的拟人化体模的头部(佛罗里达大学,盖恩斯维尔分校),我们用三种常用的扫描仪(GE Discovery CT750 HD,Siemens Definition,Philips Brilliance)测量了剂量64)在两个不同的临床站点(沃尔特·里德国家军事医学中心,国立卫生研究院)。扫描仪被设置为以与临床上用于典型头部扫描的相同的数据采集和图像重建协议进行操作,每种扫描仪的每个设施的实践各不相同。我们还使用相应的协议分析了ACR CT认证体模的图像。尽管Siemens Definition和Philips Brilliance协议仅使用常规的过滤后投影(FBP)图像重建方法,但GE Discovery还采用了其特定版本的自适应统计迭代重建(ASIR)算法,可以按所需比例进行混合使用FBP算法。我们进行了客观的图像指标分析,以评估通过FBP重建的图像的调制传递函数(MTF),噪声功率谱(NPS)和CNR。对于使用ASIR重建的图像,我们仅分析了CNR,因为预期MTF和NPS结果取决于迭代重建算法的对象。结果:OSLD测量结果显示,西门子定义扫描仪和飞利浦华晨扫描仪(分别位于两个不同的临床设施)在组织中产生的平均吸收剂量分别为42.6和43.1 mGy。当在类似的操作和图像重建条件下运行时,即没有电子管电流调制和ASIR时,GE Discovery可以提供大约相同剂量的剂量(43.7 mGy)。图像度量分析同样显示,当三个扫描仪以相似的设置运行时,与重构图像关联的MTF,NPS和CNR相互可比,并且差异可归因于所应用的重构滤镜的不同边缘增强特性。此外,当GE扫描仪使用该设施的扫描仪设置进行常规的头部检查时(应用50%ASIR,并且仅使用100%-FBP剂量的一半),图像的CNR没有明显变化。即使仅靠CNR不足以表征图像质量并证明任何减少剂量的要求,它也可以用作恒定性测试指标。结论:这项工作提出了一种直接的方法,可以将CT剂量的直接测量与客观图像指标(例如高对比度分辨率,噪声和CNR)联系起来。它证明了在拟人化头部模型中的OSLD测量可以对典型的CT头部扫描过程中所递送的组织中剂量的大小和空间分布进行现实且局部的精确估计。 ACR认证体模图像的其他客观分析可用于将测得的剂量与高对比度分辨率,噪声和CNR相关联。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号