首页> 外文期刊>Medical Physics >CT head-scan dosimetry in an anthropomorphic phantom and associated measurement of ACR accreditation-phantom imaging metrics under clinically representative scan conditions
【24h】

CT head-scan dosimetry in an anthropomorphic phantom and associated measurement of ACR accreditation-phantom imaging metrics under clinically representative scan conditions

机译:CT头扫描剂量在临床代表性扫描条件下的拟人偶像和ACR认证 - 幻影成像度量的相关测量

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To measure radiation absorbed dose and its distribution in an anthropomorphic head phantom under clinically representative scan conditions in three widely used computed tomography (CT) scanners, and to relate those dose values to metrics such as high-contrast resolution, noise, and contrast-to-noise ratio (CNR) in the American College of Radiology CT accreditation phantom. Methods: By inserting optically stimulated luminescence dosimeters (OSLDs) in the head of an anthropomorphic phantom specially developed for CT dosimetry (University of Florida, Gainesville), we measured dose with three commonly used scanners (GE Discovery CT750 HD, Siemens Definition, Philips Brilliance 64) at two different clinical sites (Walter Reed National Military Medical Center, National Institutes of Health). The scanners were set to operate with the same data-acquisition and image-reconstruction protocols as used clinically for typical head scans, respective of the practices of each facility for each scanner. We also analyzed images of the ACR CT accreditation phantom with the corresponding protocols. While the Siemens Definition and the Philips Brilliance protocols utilized only conventional, filtered back-projection (FBP) image-reconstruction methods, the GE Discovery also employed its particular version of an adaptive statistical iterative reconstruction (ASIR) algorithm that can be blended in desired proportions with the FBP algorithm. We did an objective image-metrics analysis evaluating the modulation transfer function (MTF), noise power spectrum (NPS), and CNR for images reconstructed with FBP. For images reconstructed with ASIR, we only analyzed the CNR, since MTF and NPS results are expected to depend on the object for iterative reconstruction algorithms. Results: The OSLD measurements showed that the Siemens Definition and the Philips Brilliance scanners (located at two different clinical facilities) yield average absorbed doses in tissue of 42.6 and 43.1 mGy, respectively. The GE Discovery delivers about the same amount of dose (43.7 mGy) when run under similar operating and image-reconstruction conditions, i.e., without tube current modulation and ASIR. The image-metrics analysis likewise showed that the MTF, NPS, and CNR associated with the reconstructed images are mutually comparable when the three scanners are run with similar settings, and differences can be attributed to different edge-enhancement properties of the applied reconstruction filters. Moreover, when the GE scanner was operated with the facility's scanner settings for routine head exams, which apply 50% ASIR and use only approximately half of the 100%-FBP dose, the CNR of the images showed no significant change. Even though the CNR alone is not sufficient to characterize the image quality and justify any dose reduction claims, it can be useful as a constancy test metric. Conclusions: This work presents a straightforward method to connect direct measurements of CT dose with objective image metrics such as high-contrast resolution, noise, and CNR. It demonstrates that OSLD measurements in an anthropomorphic head phantom allow a realistic and locally precise estimation of magnitude and spatial distribution of dose in tissue delivered during a typical CT head scan. Additional objective analysis of the images of the ACR accreditation phantom can be used to relate the measured doses to high contrast resolution, noise, and CNR.
机译:目的:测量辐射吸收剂量及其在三个广泛使用的计算机断层扫描仪(CT)扫描仪中的临床代表性扫描条件下的拟方向脉动的分布,并将这些剂量值与诸如高对比度分辨率,噪声和对比度的指标相关联美国放射学CT认证幻像中的噪声比(CNR)。方法:通过将光学刺激的发光剂量计(OSLD)插入专门为CT Dosimetry(佛罗里达大学,Gainesville)的拟人幻影头部,我们用三种常用的扫描仪测量剂量(GE Discovery CT750 HD,Siemens定义,飞利浦的辉煌64)在两个不同的临床网站(沃尔特芦苇国家军事医疗中心,国家卫生研究院)。扫描仪被设置为在临床上使用相同的数据获取和图像重建协议,用于典型的头部扫描,各自为每个扫描仪的每个设施的实践相应。我们还通过相应的协议分析了ACR CT认证幻像的图像。虽然Siemens定义和飞利浦的光彩协议仅使用常规过滤的后投影(FBP)图像重建方法,但GE发现还采用了其特定版本的自适应统计迭代重建(ASIR)算法,其可以以所需的比例混合使用FBP算法。我们做了一个客观的图像指标分析,评估调制传递函数(MTF),噪声功率谱(NPS)和CNR,用于用FBP重建的图像。对于使用ASIR重建的图像,我们仅分析了CNR,因为预期MTF和NPS结果将取决于迭代重建算法的对象。结果:OSLD测量表明,西门子定义和飞利浦的光彩扫描仪(位于两种不同的临床设施)分别在42.6和43.1米的组织中产生平均吸收剂量。在类似的操作和图像重建条件下运行,即没有管电流调制和ASIL,GE发现在运行时提供相同量的剂量(43.7 MGO)。图像测量分析同样显示与重建图像相关联的MTF,NP和CNR在使用类似设置运行时相互比较,并且差异可以归因于所应用的重建滤波器的不同边缘增强特性。此外,当GE扫描仪使用设施的扫描仪设置进行常规头部检查时,施加50%ASIR并仅使用大约100%-FBP剂量的大约一半,图像的CNR显示没有显着的变化。即使单独的CNR不足以表征图像质量并证明任何剂量还原权利要求,它也可用作恒定测试度量。结论:这项工作提出了一种直接的方法,用于将CT剂量的直接测量与客观图像度量(如高对比度分辨率,噪声和CNR)连接。它表明,拟人术头幻影中的OSLD测量允许在典型的CT头扫描期间递送的组织中剂量的幅度和空间分布允许逼真和局部精确地估计。对ACR认证幻像图像的其他目的分析可用于将测量剂量与高对比度分辨率,噪声和CNR相关联。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号