摘要

Since the NIST dose to water standard for 90Sr/90Y ophthalmic applicators was introduced, numerous sources have undergone calibration either at NIST or at the University of Wisconsin Accredited Dosimetry Calibration Laboratory (UWADCL). From 1997 to 2008, 222 of these beta-emitting sources were calibrated at the UWADCL, and prior reference source strength values were available for 149 of these sources. A survey of UWADCL ophthalmic applicator calibrations is presented here, demonstrating an average discrepancy of -19% with a standard deviation of +/- 16% between prior reference values and the NIST-traceable UWADCL absorbed dose to water calibrations. Values ranged from -49% to +42%.
机译:自从引入90Sr / 90Y眼药水的NIST剂量标准水以来,无论是在NIST还是在威斯康星大学认可的剂量学校准实验室(UWADCL)都对许多来源进行了校准。从1997年到2008年,在UWADCL校准了222个这些β发射源,并且有149个这些源的先前参考源强度值可用。本文介绍了UWADCL眼科用药标定的调查,结果表明,以前的参考值与NIST可追溯的UWADCL吸水剂量对水的标定之间的平均差异为-19%,标准差为+/- 16%。值范围从-49%到+ 42%。