This paper reports on a comparative study between the well-established test patterns for daily quality assurance (QA) of monitors of the American Association of Medical Physicists, Task Group 18 (AAPMtg18) and the Deutsches Institut fur Normung e.V (DIN), and a newly proposed variable test pattern. A characteristic of the test patterns currently used for the QA of monitors is their static nature: The same test pattern is always used. This enables a learning effect that may bias the results over time. To address this problem we have developed a variable pattern for the quality assurance of monitors (MoniQA) that allows an evaluation of contrast visibility, geometric distortion, resolution, global image quality including uniformity, and artifacts. The test pattern includes randomly generated elements intended to prevent the observer from learning the test. Examples are random characters that have to be discriminated from the background to evaluate the threshold luminance difference and variable positionsof different features in the test pattern. The newly proposed test patterns were generated and visualized on different viewing stations with a software tool developed in JAVA. In this study, we validated these patterns against the well-known AAPMtg18 and DIN test patterns on 22 monitors. The results showed that the MoniQA test can indicate the same monitor problems as the other well-known patterns and is significantly quicker to evaluate than the AAPMtg18 test patterns. The MoniQA pattern is a promising alternative for daily quality control of medical viewing stations.
机译:本文报告了在美国医学物理学家协会,第18任务组(AAPMtg18)和德国德意志诺曼精神研究所(DIN)建立的监视器日常质量保证(QA)测试模式之间的比较研究。新提出的可变测试模式。当前用于监视器QA的测试模式的一个特点是它们的静态特性:始终使用相同的测试模式。这使得学习效果可能随时间推移而产生偏差。为了解决这个问题,我们开发了一种可变的模式来保证监视器的质量(MoniQA),该模式可以评估对比度的可视性,几何失真,分辨率,包括均匀性和伪影的整体图像质量。测试模式包括随机生成的元素,旨在防止观察者学习测试。示例是必须与背景区分开来评估阈值亮度差异和测试图案中不同特征的可变位置的随机字符。通过使用JAVA开发的软件工具,可以在不同的观察站上生成并可视化新提出的测试模式。在这项研究中,我们针对22台显示器上的著名AAPMtg18和DIN测试模式验证了这些模式。结果表明,MoniQA测试可以指示与其他知名模式相同的监视器问题,并且比APMPMtg18测试模式明显更快地进行评估。 MoniQA模式是用于医疗观察站日常质量控制的有前途的替代方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号