首页> 外文期刊>Medical Physics >Absolute dose measurements by means of a small cylindrical ionization chamber for very high dose per pulse high energy electron beams.
【24h】

Absolute dose measurements by means of a small cylindrical ionization chamber for very high dose per pulse high energy electron beams.

机译:借助于小圆柱电离室的绝对剂量测量,用于每脉冲高剂量高能电子束。

获取原文
获取原文并翻译 | 示例
           

摘要

Very high dose per pulse (3-13 cGy/pulse) high energy electron beams are currently produced by special linear accelerators (linac) dedicated to Intra Operative Radiation Therapy (IORT). The electron beams produced by such linacs are collimated by special Perspex applicators of various size and cylindrically shaped. The biggest problems from the dosimetric point of view are caused by the high dose-per-pulse values and the use of inclined applicators. In this work measurements of absolute dose for the inclined applicators were done by using a small cylindrical ionization chamber, type CC01 (Wellhofer), a parallel plane ionization chamber type Markus (PTW 23343) and radiochromic films type EBT. We show a method which allows calculating the quality correction factors for CC01 chamber with an uncertainty of 1% and the absolute dose value for the inclined applicators using CC01 with an uncertainty of 3.1% for electron beams of energy of 6 and 7 MeV produced by the linac dedicated to IORT Novac7.
机译:目前,专用于手术中放射治疗(IORT)的特殊线性加速器(直线加速器)产生了非常高的每脉冲剂量(3-13 cGy /脉冲)高能电子束。由这种直线加速器产生的电子束由各种尺寸和圆柱形的特殊有机玻璃施加器准直。从剂量学的角度来看,最大的问题是每脉冲高剂量值和倾斜施药器的使用。在这项工作中,通过使用CC01型(Wellhofer),圆柱形Markus型平行电离室(PTW 23343)和EBT型放射致变色薄膜的小型圆柱形电离室来测量倾斜施加器的绝对剂量。我们展示了一种方法,该方法可以计算不确定度为1%的CC01腔室的质量校正因子,以及使用由CC01产生的能量为6和7 MeV的电子束的不确定度为3.1%的CC01来计算倾斜施加器的绝对剂量值。专门用于IORT Novac7的直线加速器。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号