Characteristics of optically stimulated luminescence dosimeters in the spread-out Bragg peak region of clinical proton beams.

机译:临床质子束扩散布拉格峰区域中的光激发发光剂量计的特征。

获取原文
获取原文并翻译 | 示例

摘要

Optically stimulated luminescent detectors (OSLDs) have a number of advantages in radiation dosimetry making them excellent dosimeters for quality assurance and patient dose verification. Although the dosimeters have been investigated in several modalities, relatively little work has been done in examining the dosimeters for use in clinical proton beams. This study examined a number of characteristics of the response of the dosimeters in the spread-out Bragg peak (SOBP) region of clinical proton beams.Optically stimulated luminescence (OSL) dosimeters from Landauer, Inc., specifically the nanoDot dosimeter, were investigated. These dosimeters were placed in a special phantom with a recess to fit the dosimeters without an air gap. Beams with nominal energies of 160, 200, and 250 MeV were used in the passively-scattered proton beam at the MD Anderson Cancer Center Proton Therapy Center. Dosimetric properties including linearity, field size dependence, energy dependence, residual signal as a function of cumulative dose, and postirradiation fading were investigated by taking measurements at the center of SOBPs.The dosimeters showed 1% supralinearity at 200 cGy and 5% supralinearity at 1000 cGy. No noticeable field size dependence of the detector was found for field sizes from 2?×?2 cm(2) to 18?×?18 cm(2). Residual signal as a function of cumulative dose showed a small increase for measurements up to 1000 cGy. Readout signal depletion of the dosimeters after consecutive readings showed a slightly larger depletion in protons for doses up to 500 cGy but not by a clinically significant amount. Within the center of various SOBP widths and proton energies the variation in response was less than 2%. An average beam quality factor of 1.089 with experimental standard deviation of 0.007 was determined and applied to the data such that the results were within 1.2% of ion chamber data.The nanoDot OSL dosimeter characteristics were studied in the SOBP region of clinical proton beams. To achieve accurate dosimetric readings, corrections to the dosimeter response were applied. Corrections tended to be minimal or broadly consistent. The nanoDot OSLD was found to be an acceptable dosimeter for measurement in the SOBP region for a range of clinical proton beams.
机译:光激发发光检测器(OSLD)在辐射剂量测定中具有许多优势,使其成为用于质量保证和患者剂量验证的优秀剂量计。尽管已经以几种方式对剂量计进行了研究,但是在检查用于临床质子束的剂量计中所做的工作相对较少。这项研究检查了临床质子束扩散布拉格峰(SOBP)区域中剂量计响应的许多特征,并研究了Landauer,Inc.的光激发发光(OSL)剂量计,特别是nanoDot剂量计。将这些剂量计放置在带有凹口的特殊体模中,以使其适合而没有气隙。 MD安德森癌症中心质子治疗中心的被动散射质子束中使用了标称能量分别为160、200和250 MeV的束。通过在SOBPs的中心进行测量,研究了线性,剂量场依赖性,能量依赖性,残余信号随累积剂量的函数以及辐照后衰落等剂量学特性。在200 cGy处,剂量计显示1%的超线性,在1000 cGy处显示5%的超线性cGy。对于2?×?2 cm(2)到18?×?18 cm(2)的场大小,没有发现检测器的场大小依赖性。残留信号作为累积剂量的函数,对于高达1000 cGy的测量显示出很小的增加。连续读数后剂量计的读出信号消耗显示,对于剂量高达500 cGy的质子,消耗的质子略大,但临床上没有显着量。在各种SOBP宽度和质子能量的中心内,响应变化小于2%。确定了平均束质量因数1.089和实验标准偏差0.007,并将其应用于数据,以使结果在离子室数据的1.2%范围内。在临床质子束的SOBP区域中研究了nanoDot OSL剂量计特性。为了获得准确的剂量学读数,对剂量计响应进行了校正。校正趋向于极少或基本一致。发现nanoDot OSLD是用于一系列临床质子束在SOBP区域测量的可接受剂量计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号