Reduction of false arterial blood pressure alarms using signal quality assessment and relationships between the electrocardiogram and arterial blood pressure.

机译:使用信号质量评估以及心电图和动脉血压之间的关系来减少错误的动脉血压警报。

获取原文
获取原文并翻译 | 示例
           

摘要

The paper presents an algorithm for reducing false alarms related to changes in arterial blood pressure (ABP) in intensive care unit (ICU) monitoring. The algorithm assesses the ABP signal quality, analyses the relationship between the electrocardiogram and ABP using a fuzzy logic approach and post-processes (accepts or rejects) ABP alarms produced by a commercial monitor. The algorithm was developed and evaluated using unrelated sets of data from the MIMIC database. By rejecting 98.2% (159 of 162) of the false ABP alarms produced by the monitor using the test set of data, the algorithm was able to reduce the false ABP alarm rate from 26.8% to 0.5% of ABP alarms, while accepting 99.8% (441 of 442) of true ABP alarms. The results show that the algorithm is effective and practical, and its use in future patient monitoring systems is feasible.
机译:本文提出了一种算法,用于减少重症监护病房(ICU)监测中与动脉血压(ABP)变化有关的虚假警报。该算法评估ABP信号质量,使用模糊逻辑方法分析心电图与ABP之间的关系,并对商用监视器产生的ABP警报进行后处理(接受或拒绝)。该算法是使用MIMIC数据库中不相关的数据集开发和评估的。通过拒绝使用测试数据集的监视器生成的98.2%(162个中的159个)错误ABP警报,该算法能够将ABP错误警报率从ABP警报的26.8%降低到0.5%,同时接受99.8% (442之441)的真实ABP警报。结果表明该算法是有效和实用的,在未来的病人监护系统中使用是可行的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号