首页> 外文期刊>Magnetic resonance in medicine: official journal of the Society of Magnetic Resonance in Medicine >Quantitative evaluation of susceptibility and shielding effects of nitinol, platinum, cobalt-alloy, and stainless steel stents.
【24h】

Quantitative evaluation of susceptibility and shielding effects of nitinol, platinum, cobalt-alloy, and stainless steel stents.

机译:镍钛合金,铂,钴合金和不锈钢支架的敏感性和屏蔽效果的定量评估。

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this study is to quantitatively estimate the shielding and susceptibility effects of commonly used metallic stents on MR signal. Two experiments were performed using a 3D gradient echo sequence with short TE to image a stent phantom: 1) short TR and high flip angle (contrast enhanced MRA parameters), and 2) long TR (TR T(1)) and low flip angle. The factor characterizing susceptibility effects was estimated from the signal phase of the first experiment, and then the factor characterizing the shielding effects was derived from the second experiment. Susceptibility induced signal loss was negligible (<1%) for nonstainless-steel (nitinol, platinum, and cobalt-alloy) stents and totally destructive (100%) for the stainless steel stent. Signal loss due to RF shielding was 31-62% for nitinol stents, 14-50% for platinum stents, 50-77% for the cobalt-alloy stents (undetermined for the stainless steel stent), varied with stent orientation, diameter, and wall geometry. In summary, stents made of nitinol, platinum, and cobalt-alloy have negligible susceptibility effects but stents made of stainless steel may have complete dephasing. All stents have substantial shielding effects, which vary with composition, geometry, and orientation. Large platinum stents may have the smallest artifacts and are the best suited for postinterventional MR imaging. Magn Reson Med 49:972-976, 2003.
机译:这项研究的目的是定量评估常用金属支架对MR信号的屏蔽和敏感性。使用具有短TE的3D梯度回波序列进行了两个实验,以对支架体模进行成像:1)短TR和高翻转角(对比增强的MRA参数),以及2)长TR(TR T(1))和低TR翻转角度。从第一个实验的信号相位估计表征磁敏效应的因子,然后从第二个实验得出表征屏蔽效应的因子。对于非不锈钢(镍钛合金,铂和钴合金)支架,易感性导致的信号损失可忽略不计(<1%),而对于不锈钢支架,则完全具有破坏性(100%)。镍钛合金支架的射频屏蔽信号损失为31-62%,铂合金支架为14-50%,钴合金支架(不锈钢支架未确定)为50-77%,随支架方向,直径和墙的几何形状。总之,由镍钛合金,铂和钴合金制成的支架对磁化的影响可忽略不计,但由不锈钢制成的支架可能会完全脱相。所有支架均具有明显的屏蔽效果,该效果随成分,几何形状和方向而变化。大型铂金支架可能具有最小的伪影,并且最适合介入后MR成像。 Magn Reson Med 49:972-976,2003年。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号