High-resolution blood flow velocity measurements in the human finger.

机译:人手指中的高分辨率血流速度测量。

获取原文
获取原文并翻译 | 示例

摘要

MR phase contrast blood flow velocity measurements in the human index finger were performed with triggered, nontriggered, and cine acquisition schemes. A strong (G(max) = 200 mT/m), small bore (inner diameter 12 cm) gradient system inserted in a whole body 3 Tesla MR scanner allowed high-resolution imaging at short echo times, which decreases partial volume effects and flow artifacts. Arterial blood flow velocities ranging from 4.9-19 cm/sec were measured, while venous blood flow was significantly slower at 1.5-7.1 cm/sec. Taking into account the corresponding vessel diameters ranging from 800 microm to 1.8 mm, blood flow rates of 3.0-26 ml/min in arteries and 1.2-4.8 ml/min in veins are obtained. The results were compared to ultrasound measurements, resulting in comparable blood flow velocities in the same subjects. Magn Reson Med 45:716-719, 2001. Copyright 2001 Wiley-Liss, Inc.
机译:人类食指中的MR相衬血流速度测量是通过触发,非触发和电影采集方案进行的。强大的(G(max)= 200 mT / m),小孔径(内径12 cm)梯度系统插入全身3 Tesla MR扫描仪,可在短回波时间内进行高分辨率成像,从而减少了部分体积效应和流量文物。测得的动脉血流速度范围为4.9-19 cm / sec,而静脉血流的速度明显慢于1.5-7.1 cm / sec。考虑到相应的血管直径为800微米至1.8毫米,在动脉中的血流速度为3.0-26 ml / min,在静脉中的血流速度为1.2-4.8 ml / min。将结果与超声测量结果进行比较,得出相同受试者的血流速度相当。 Magn Reson Med 45:716-719,2001。版权所有2001 Wiley-Liss,Inc.。