A flow-through optosensor has been prepared for the sensitive and selective determination of pyridoxine (vitamin B_6) in aqueous solutions. The sensor was developed in conjunction with a monochannel flow-in-jection analysis system with fluorimetric detection using Sephadex SP-C25 resin as an active sorbent substrate. This method of determination is carried out without any derivatization. The wavelengths of excitation and emission were 295 and 385 nm, respectively. When a HCl (10~(-3) mol L~(-1))/NaCl (3 * 10~(-2) mol L~(-1)) solution is used as carrier solution, the sensor responds linearly in the measuring range of 5-200, 10-400 and 50-1800 ng mL~(-1) with detection limits of 0.33, 0.67, and 5.70 ng mL~(-1) for 2000, 1000 and 200 #mu#L of sample volume, respectively. The relative standard deviation for ten independent determinations is less than 0.75% for 0.2 and 1.0 mL of sample volumes used, and 1.31% for 2.0 mL of sample volume used. the method was satisfactorily applied to the determination of vitamin B_6 in pharmaceutical preparations.
机译:已经准备了一种流通型光电传感器,用于灵敏和选择性地测定水溶液中的吡selective醇(维生素B_6)。该传感器与单通道流动注射分析系统一起开发,该系统具有荧光检测功能,使用Sephadex SP-C25树脂作为活性吸附剂基质。这种确定方法无需任何衍生即可进行。激发和发射的波长分别为295和385 nm。当使用HCl(10〜(-3)mol L〜(-1))/ NaCl(3 * 10〜(-2)mol L〜(-1))溶液作为载体溶液时,传感器在测量范围为5-200、10-400和50-1800 ng mL〜(-1),对2000、1000和200#mu#L的样品的检出限为0.33、0.67和5.70 ng mL〜(-1)音量。十次独立测定的相对标准偏差对于使用0.2和1.0 mL样品体积小于0.75%,对于使用2.0 mL样品体积小于1.31%。该方法可令人满意地用于药物制剂中维生素B_6的测定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号