首页> 外文期刊>European journal of trauma and emergency surgery: official publication of the European Trauma Society >Rupture of Flexor Pollicis Longus Tendon: A Complication of Volar Locking Plating of the Distal Radius
【24h】

Rupture of Flexor Pollicis Longus Tendon: A Complication of Volar Locking Plating of the Distal Radius

机译:屈肌肌腱断裂:远端Rad掌锁定钢板的并发症

获取原文
获取原文并翻译 | 示例

摘要

We report an unusual case of complete rupture of the flexor pollicis longus tendon following volar locking plating for a distal radius fracture. We believe that the prominence of a distal locking screw head predisposed to the rupture of the tendon. We highlight that correctly attaching the distal locking screws to the plate is essential for obtaining the correct biomechanics of the device and preventing flexor tendon rupture.
机译:我们报告了一个罕见的病例,在vol骨锁定钢板治疗radius骨远端骨折后,屈曲的耻骨长肌腱完全破裂。我们认为,远端锁定螺钉头突出会导致肌腱断裂。我们强调指出,正确地将远端锁定螺钉连接到板上对于获得正确的器械生物力学和防止屈肌腱断裂至关重要。