Investigation of capacitor failures in an automated external defibrillator.

机译:调查自动体外除颤器中的电容器故障。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Over the past 2 years, the Canadian Medical Devices Bureau has received a number of reports of capacitor failures on the high voltage board of an automated external defibrillator. Twenty-five cases of broken capacitor leads were found during routine preventive maintenance by the biomedical engineering staff of the institutions reporting the incidents. The Bureau has carried out a laboratory investigation to determine the effect of missing capacitors on the energy delivered by the defibrillator and to assess whether these capacitor failures represent a significant risk to patients. Our findings indicate that the automated external defibrillator will not perform acceptably with two broken capacitors. They further suggest that, during preventive maintenance, operators should use a defibrillator analyser to measure the delivered energy output rather than using the internal energy measurement circuit within the automated external defibrillator.
机译:在过去的两年中,加拿大医疗器械局收到了许多有关自动体外除颤器高压板上电容器故障的报告。报告事件的机构的生物医学工程人员在例行的预防性维护期间,发现了25例电容器引线断裂的情况。该局已进行了一项实验室调查,以确定缺少电容器对除颤器输送的能量的影响,并评估这些电容器故障是否对患者构成重大风险。我们的发现表明,自动外部除颤器在两个坏掉的电容器上不能令人满意地工作。他们进一步建议,在预防性维护期间,操作员应使用除颤器分析仪来测量输出的能量输出,而不是使用自动外部除颤器中的内部能量测量电路。