Ventilator assessment of respiratory mechanics in paediatric intensive care.

机译:儿科重症监护室呼吸机的呼吸机评估。

获取原文
获取原文并翻译 | 示例
           

摘要

Many modern "paediatric" mechanical ventilators have in-built features for estimation of respiratory mechanics which could be useful in the management of ventilated infants and children. The aim of this study was to determine if such measurements were reproducible and accurate. Ventilator (Draeger Evita 4) displayed compliance (Cvent) and resistance (Rvent) values were assessed and compared to the results of respiratory system mechanics (respiratory system compliance (Crs) and resistance (Rrs)) measurements obtained using a single breath occlusion technique. Seventeen children (median age 5.1; range 0.3 to 16 yrs) were studied on 24 occasions. The mean coefficients of variations for the techniques were similar (Cvent 13%; Crs 11%; Rvent 16%; Rrs 14%). The mean (SD) Crs (22.8 (12.3) ml/cmH2O) did not differ significantly from Cvent (22.1 (12.7) ml/cm H2O) but the mean Rrs 21.0 (12.7) cmH2O/l/s was significantly higher than the mean Rvent 32.0 (32.0) cmH2O/l/s (p = 0.03). Bland and Altman analysis demonstrated a mean difference of -10.94 cmH2O/l/s (SD 24.1) between Rrs and Rvent; the agreement between Rrs and Rvent decreased as Rrs increased (p = 0.008). CONCLUSIONS: Ventilator assessment of compliance, but not resistance, using the Evita 4 is reproducible and reliable.
机译:许多现代的“儿科”机械呼吸机具有用于估计呼吸力学的内置功能,这些功能可能对管理通风的婴幼儿有用。这项研究的目的是确定这种测量是否可重复和准确。评估呼吸机(Draeger Evita 4)的顺应性(Cvent)和阻力(Rvent)值,并将其与使用单次呼吸阻塞技术获得的呼吸系统力学结果(呼吸系统顺应性(Crs)和阻力(Rrs))进行比较。对17个儿童(中位年龄5.1;范围0.3至16岁)进行了24次研究。该技术的平均变异系数相似(Cvent 13%; Crs 11%; Rvent 16%; Rrs 14%)。平均(SD)Crs(22.8(12.3)ml / cmH2O)与Cvent(22.1(12.7)ml / cm H2O)没有显着差异,但平均Rrs 21.0(12.7)cmH2O / l / s显着高于平均值Rvent 32.0(32.0)cmH2O / l / s(p = 0.03)。 Bland和Altman分析显示Rrs和Rvent之间的平均差异为-10.94 cmH2O / l / s(SD 24.1); Rrs和Rvent之间的一致性随Rrs的增加而降低(p = 0.008)。结论:使用Evita 4对呼吸机进行依从性评估,而不是阻力评估,可再现且可靠。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号