首页> 外文期刊>Epilepsy research >Usefulness of a 1.5 T MRI-compatible EEG electrode system for routine use in the intensive care unit of a tertiary care hospital.
【24h】

Usefulness of a 1.5 T MRI-compatible EEG electrode system for routine use in the intensive care unit of a tertiary care hospital.

机译:1.5 T MRI兼容的EEG电极系统在三级护理医院的重症监护室常规使用的有用性。

获取原文
获取原文并翻译 | 示例
       

摘要

BACKGROUND: Continuous electroencephalogram (cEEG) recordings are being increasingly used in intensive care units (ICUs) to detect epileptic seizures and other changes. MRI scans can interrupt such recordings if the EEG electrodes need to be removed and important data can be missed. METHODS: We retrospectively examined EEG recordings from ICU patients who underwent MRI scans, comparing those from patients with the MRI-compatible EEG electrodes with those who had to have the EEG electrodes removed before scanning. We also examined technical aspects of the recording and scalp abrasions in both groups. RESULTS: Fourteen of 31 (45%) EEG recordings with the MRI-compatible electrode system in patients that underwent MRI scans between 03:00 p.m. and 07:00 a.m. (when technologists were not available) captured seizures. In contrast, all of the six EEG recordings with the MRI-incompatible electrode system in patients that underwent MRI scanning during the same interval were interrupted and had no data. Recordingcharacteristics of the EEGs were comparable between the two groups and none had scalp abrasions. CONCLUSION: A significant proportion of patients undergoing MRI scans with the MRI-compatible EEG electrodes had seizures that would have been missed if the MRI-incompatible EEG electrodes had been used.
机译:背景:重症监护病房(ICU)越来越多地使用连续脑电图(cEEG)记录来检测癫痫发作和其他变化。如果需要移除EEG电极并且可能遗漏重要数据,则MRI扫描会中断此类记录。方法:我们回顾性检查了接受过MRI扫描的ICU患者的EEG记录,将那些与MRI兼容的EEG电极的患者与那些必须在扫描前取下EEG电极的患者进行了比较。我们还研究了两组记录和头皮擦伤的技术方面。结果:在下午03:00之间接受MRI扫描的患者中,使用MRI兼容电极系统的31例脑电图记录中有14例(占45%)。凌晨07:00(技术人员不在时)捕获了癫痫发作。相反,在相同时间间隔内进行MRI扫描的患者使用MRI不兼容电极系统进行的6次EEG记录均被中断,没有数据。两组的脑电图记录特征相当,并且没有头皮擦伤。结论:很大一部分接受MRI兼容的EEG电极进行MRI扫描的患者,如果使用了MRI兼容的EEG电极,将会错过癫痫发作。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号