Macular thickness measurements in healthy eyes using six different optical coherence tomography instruments.

机译:使用六种不同的光学相干断层扫描仪器在健康眼睛中进行黄斑厚度测量。

获取原文
获取原文并翻译 | 示例

摘要

PURPOSE: To compare central retinal thickness (CRT) measurements in healthy eyes by different commercially available OCT instruments and to compare the intersession reproducibility of such measurements. METHODS: Six different OCT instruments (Stratus OCT [Carl Zeiss Meditec, Inc. Dublin, CA], SOCT Copernicus [Reichert/Optopol Technology, Inc., Depew, NY], Spectral OCT/SLO [Opko/OTI, Inc., Miami, FL], RTVue-100 [Optovue Corp., Fremont, CA], Spectralis HRA+OCT [Heidelberg Engineering, Inc., Heidelberg, Germany], and Cirrus HD-OCT [Carl Zeiss Meditec, Inc.]) were used to assess CRT in both eyes of healthy subjects. Measurements were performed in two different sessions on the same day with each of the systems. From these measurements, the mean CRT was calculated. For the assessment of the intersession reproducibility of the instruments, we calculated the coefficient of the variation of test-retest variation. RESULTS: Twenty healthy subjects were included in the study. Compared with the Stratus OCT all spectral OCT instruments showed significantly higher CRTs. The Spectralis HRA+OCT and Cirrus HD-OCT showed similar CRT values but significantly higher values than did all other instruments. The coefficients of variation for repeated measurements was 3.33% for the Stratus OCT, 0.46% for the Spectralis HRA+OCT, 3.09% for the Cirrus HD-OCT, 2.23% for the OCT/SLO, 2.77% for the RTVue-100 OCT, and for the SOCT 3.5%, respectively. discussion. The six OCT systems provided different results for CRT. The measurements with the Stratus OCT showed the lowest thicknesses, whereas those with the Cirrus HD-OCT and Spectralis HRA+OCT yielded the highest ones. These discrepancies can be explained by the differences in the retinal segmentation algorithms used by the various OCT systems. Whereas the Spectralis HRA+OCT and Cirrus HD-OCT include the RPE layer in the retinal segmentation, the other instruments do not. The data imply that the different OCT systems cannot be used interchangeably for the measurement of macular thickness.
机译:目的:通过不同的市售OCT仪器比较健康眼睛中中央视网膜厚度(CRT)的测量值,并比较这些测量值在闭会期间的再现性。方法:六种不同的OCT仪器(Stratus OCT [Carl Zeiss Meditec,Inc. Inc.,加利福尼亚州都柏林],SOCT Copernicus [Reichert / Optopol Technology,Inc.,纽约州Depew],光谱OCT / SLO [Opko / OTI Inc.,迈阿密,FL],RTVue-100 [Optovue Corp.,Fremont,CA],Spectralis HRA + OCT [Heidelberg Engineering,Inc.,Heidelberg,Germany]和Cirrus HD-OCT [Carl Zeiss Meditec,Inc.])用于在健康受试者的双眼中评估CRT。每个系统在同一天的两个不同会话中进行测量。根据这些测量结果,计算出平均CRT。为了评估仪器的会话间可重复性,我们计算了重测变异的变异系数。结果:20名健康受试者被纳入研究。与Stratus OCT相比,所有光谱OCT仪器均显示出显着更高的CRT。 Spectralis HRA + OCT和Cirrus HD-OCT显示出相似的CRT值,但其值明显高于所有其他仪器。对于Stratus OCT,重复测量的变异系数为3.33%,Spectralis HRA + OCT为0.46%,Cirrus HD-OCT为3.09%,OCT / SLO为2.23%,RTVue-100 OCT为2.77%,和SOCT分别为3.5%。讨论。六个OCT系统为CRT提供了不同的结果。用Stratus OCT进行的测量显示出最小的厚度,而使用Cirrus HD-OCT和Spectralis HRA + OCT进行的测量得到了最大的厚度。这些差异可以通过各种OCT系统使用的视网膜分割算法的差异来解释。 Spectralis HRA + OCT和Cirrus HD-OCT在视网膜分割中包括RPE层,而其他仪器则没有。数据表明,不同的OCT系统不能互换地用于黄斑厚度的测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号