Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer.

机译:角膜参数对使用脉动眼血流眼压计和Goldmann压平眼压计进行测量的影响。

获取原文
获取原文并翻译 | 示例
       

摘要

AIMS: To investigate the effect of central corneal thickness and corneal curvature on intraocular pressure measurements using the pulsatile ocular blood flow tonograph and the Goldmann applanation tonometer, and to assess the agreement between the pulsatile ocular blood flow tonograph and the Goldmann applanation tonometer in intraocular pressure measurement. METHODS: 479 subjects underwent intraocular pressure measurements with the Goldmann applanation tonometer and the pulsatile ocular blood flow tonograph. Of these, 334 patients underwent additional measurement of central corneal thickness with an ultrasonic pachymeter and corneal curvature measurement with a keratometer. RESULTS: The intraocular pressure measurements obtained with both the Goldmann applanation tonometer and the pulsatile ocular blood flow tonograph varied with central corneal thickness and mean keratometric reading. Intraocular pressure measured using the Goldmann applanation tonometer increased by 0.027 mm Hg per micro m increase in central corneal thickness. Intraocular pressure measured using the pulsatile ocular blood flow tonograph increased by 0.048 mm Hg per micro m increase in central corneal thickness. For an increase of 1 mm of mean corneal curvature there was rise in intraocular pressure of 1.14 mm Hg measured by the Goldmann applanation tonometer and of 2.6 mm Hg measured by the pulsatile ocular blood flow tonograph. When compared to the Goldmann applanation tonometer, the pulsatile ocular blood flow tonograph underestimated at low intraocular pressure and overestimated at higher intraocular pressure. CONCLUSION: Central corneal thickness and corneal curvature affected measurements obtained with the pulsatile ocular blood flow tonograph more than they affected measurements obtained with the Goldmann applanation tonometer.
机译:目的:研究使用脉动眼血流计和戈德曼压平眼压计测量中央角膜厚度和角膜曲率对眼内压的影响,并评估脉动眼血流血压计和戈德曼压平眼压计在眼压中的一致性测量。方法:对479名受试者进行了戈德曼压平眼压计和脉动眼血流眼压计的眼压测量。其中,334例患者接受了超声测厚仪的额外角膜中央厚度测量和角膜曲率计的角膜曲率测量。结果:用戈德曼压平眼压计和脉动眼血流眼压计获得的眼内压测量值随中央角膜厚度和平均角膜曲率读数而变化。使用戈德曼压平眼压计测量的眼内压每增加1毫米汞柱,中央角膜厚度增加0.027毫米汞柱。使用脉动眼血流眼压计测量的眼内压每增加1毫米汞柱,中央角膜厚度增加0.048毫米汞柱。对于增加1mm的平均角膜曲率,通过Goldmann压平眼压计测量的眼内压升高为1.14mmHg,通过搏动性眼血流眼压计测量的眼内压升高为2.6mmHg。与高曼压平眼压计相比,脉动眼血流眼压计在低眼压下低估而在高眼压下高估。结论:中央角膜厚度和角膜曲率对使用脉动眼血流眼压计进行测量的影响要大于对戈德曼压平眼压计进行测量的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号