Effects of central corneal thickness and corneal curvature on the intraocular pressure measurement by Goldmann applanation tonometer and ocular blood flow pneumatonometer.

机译:中心角膜厚度和角膜曲率对戈德曼压平眼压计和眼血流量肺动脉压计测量眼内压的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To assess the effects of central corneal thickness (CCT) and corneal curvature (CC) on the measurements of intraocular pressure (IOP) using Goldmann applanation tonometer (GAT) and the ocular blood flow pneumatonometer (OBFT). METHODS: 104 patients were recruited from a glaucoma clinic. The CCT was measured using ultrasound pachymetry and the mean radius of CC using a keratometer. The IOP of each eye was measured using both GAT and the OBFT in a random order. Right eyes only were analysed for statistical purposes. RESULTS: The mean (+/-SD) IOP by GAT and OBFT was 18.2 mmHg (+/-4.4) and 18.2 mmHg (+/-4.0), respectively, with no statistically significant difference. IOP measurement with both instruments varied with CCT and CC. GAT showed an IOP increase of 0.40 mmHg per 10 microm increase of CCT and OBFT showed an increase of 0.38 mmHg in IOP per 10 microm increase of CCT. Multiple regression analysis showed that the effect of CCT was statistically significant (P<0.001) on IOP recorded by both the GAT and OBFT but CC did not have a statistically significant effect on IOP recordings performed by either technique. CONCLUSION: IOP measurements by GAT and OBFT are positively correlated with CCT with both tonometers being similarly affected. There was no significant correlation between CC and IOP measured by either tonometer.
机译:目的:评估使用戈德曼压平眼压计(GAT)和眼血流量肺动脉压计(OBFT)评估中央角膜厚度(CCT)和角膜曲率(CC)对眼内压(IOP)测量的影响。方法:从青光眼诊所招募104名患者。使用超声测厚法测量CCT,并使用角膜曲率计测量CC的平均半径。使用GAT和OBFT以随机顺序测量每只眼睛的IOP。仅出于统计目的对右眼进行了分析。结果:GAT和OBFT的平均IOP(+/- SD)分别为18.2 mmHg(+/- 4.4)和18.2 mmHg(+/- 4.0),差异无统计学意义。两种仪器的IOP测量随CCT和CC的不同而不同。 GAT显示每增加10微米CCT,IOP会增加0.40 mmHg,OBFT显示每增加10微米CCT,IOP会增加0.38 mmHg。多元回归分析表明,CCT对GAT和OBFT记录的IOP具有统计学意义(P <0.001),但CC对这两种技术进行的IOP记录均无统计学意义。结论:GAT和OBFT对IOP的测量与CCT呈正相关,两个眼压计也受到类似的影响。两种眼压计测量的CC和IOP之间均无显着相关性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号