首页> 外文期刊>Instruments and Experimental Techniques >Thermostatting of Condensed Samples in the Spectrometer when using the Attenuated Total Reflectance Method
【24h】

Thermostatting of Condensed Samples in the Spectrometer when using the Attenuated Total Reflectance Method

机译:使用衰减全反射法时光谱仪中冷凝样品的恒温

获取原文
获取原文并翻译 | 示例
           

摘要

The created external thermostatting system of condensed samples in the cell of a MIDAC M4000 infrared Fourier spectrometer is described. A thermostabilized metal bar that was worked out by the authors is the basic element of the system. The system is intended for investigating the IR spectra of samples using the attenuated total reflectance method (ATR spectroscopy method). When the ATR method is used, the standard range of thermostatting of samples in this spectrometer model is from room temperature to 100℃ with an accuracy of ±1℃. The desired temperature of the bar is set by the flow of a cooling fluid, which is supplied from an external thermostat, and it is registered with two thermometers (contact for the cell and infrared for a sample). The proposed approach allows samples to be thermostatted in the temperature range from -12 to 80℃ with an accuracy of ±0.5℃.
机译:描述了在MIDAC M4000红外傅里叶光谱仪池中创建的冷凝样品的外部恒温系统。作者设计的热稳定金属棒是系统的基本元素。该系统旨在使用衰减全反射法(ATR光谱法)研究样品的红外光谱。当使用ATR方法时,该光谱仪模型中样品的标准恒温范围是从室温到100℃,精度为±1℃。棒的所需温度由冷却液的流动来设定,该冷却液由外部恒温器提供,并通过两个温度计(用于电池的触点和用于样品的红外线)记录。所提出的方法允许将样品在-12至80℃的温度范围内恒温,精度为±0.5℃。