An approach for the performance analysis of an uncooled infrared bolometer imager

机译:一种非冷却式红外辐射热成像仪性能分析的方法

获取原文
获取原文并翻译 | 示例
           

摘要

As an approach for analyzing an uncooled infrared bolometer imager, a method is introduced whichsupposes that a thermal system is built into the infrared radiation detection mechanism. In thisanalysis, the bi-directional radiation energy flows between two thermal elements are considered anda radiation thermal conductance between them is defined. Based on the values of the radiation thermalconductances and conventional material thermal conductance, an equivalent thermal circuit for abolometer detector element is extracted. Calculations were carried out using software written in visualC++. The analysis shows a Noise Equivalent Temperature Difference (NETD) of 50 mK is obtained as atypical value, assuming that no additional noise other than temperature fluctuation noise exists, andthat the NETD becomes better as the system time constant built into the equivalent thermal circuitincreases.
机译:作为分析未冷却的红外辐射热成像仪的一种方法,介绍了一种方法,该方法假定在红外辐射检测机构中内置了一个热系统。在此分析中,考虑了两个热元件之间的双向辐射能量流,并定义了它们之间的辐射热导。基于辐射热导和常规材料热导的值,提取用于辐射热测定仪检测器元件的等效热回路。使用visualC ++编写的软件进行计算。分析表明,假设不存在除温度波动噪声以外的其他噪声,并且获得等效热电路中内置的系统时间常数,则NETD会获得50 mK的非典型噪声等效温度差(NETD),并且非典型值。