Histologic and histomorphometric report of three immediately loaded screw implants retrieved from man after a three-year loading period.

机译:三年加载期后从人体内取出的三枚立即加载的螺钉植入物的组织学和组织形态学报告。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: The aim of the present study, in man, was an histologic and histomorphometric analysis of the peri-implant tissues in 3 immediately loaded screw implants retrieved, because of fracture, after a 3 year loading period. RESULTS: Mineralized tissue was present at the interface of all implants. The mean bone to implant contact percentage for the 3 implants was 56.3% +/- 5%. The histologic and histomorphometric data showed that mineralized tissues can be obtained in immediately loaded screw implants and that these mineralized tissues are maintained over a period of 3 years. The response of the osseous tissue was not disturbed by the stresses and strains transmitted at the interface. CONCLUSION: A high primary stability, obtained by the large threads of the screw-shaped implant, was probably a key factor in the long-term maintenance of a mineralized bone interface. The primary stability helps to resist micromotion, i.e., the relative movements between the implant surface and surrounding bone during functional loading.
机译:目的:本研究的目的是,在人类中,对3年负荷期后因骨折而立即取回的3颗因骨折而立即取回的螺钉植入物的种植体周围组织进行组织学和组织形态学分析。结果:所有植入物的界面均存在矿化组织。 3种植入物的平均骨与植入物接触百分比为56.3%+/- 5%。组织学和组织形态计量学数据表明,可以在立即加载的螺钉植入物中获得矿化的组织,并且这些矿化的组织可以维持3年。在界面处传递的应力和应变不会干扰骨组织的反应。结论:螺钉状植入物的大螺纹具有很高的基本稳定性,这可能是长期维持矿化骨界面的关键因素。主要的稳定性有助于抵抗微动,即在功能性负载过程中植入物表面与周围骨骼之间的相对运动。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号